Terminal contacts
1–4
+ sense (load cells)
Sense +
white
Terminal contact
IND
+ sense (connection cable)
Terminal contacts
1–4
+ supply voltage (load cells)
red
Terminal contact
IND
+ supply voltage (connection cable)
Terminal contacts
1–4
Screen (load cells)
Screen
yellow
Terminal contact
IND
Screen (connection cable)
1
Jumper
2
Jumper
Connects the load cell supply line to the
sense line. Both jumpers must be opened
when 6-wire load cell cables are used.
4.3 Cable gland
The cable glands each have one sealing ring.
The M20 cable gland can be itted with a reduction sealing ring if cables ≤8 mm are used
(see Chapter
The M25 cable gland must be itted with a four-seal insert to be able to feed in up to four
load cell cables.
The cable wires are connected to the terminals inside the device.
NOTICE
Property damage is possible.
If a seal insert is not used, it must be sealed with a supplied locking pin.
NOTICE
Property damage is possible.
Regularly check the itted cable gland for tightness and re-tighten it, if necessary.
4 Installation and connection information
Cable Junction Box MP 90/04
Minebea Intec
EN-11