MP 26/10
Installation and Calibration Manual
Minebea Intec
EN-11
3.4.
Key Pad
There are 3 keys in the Device. They are
Enter key
Increment key
Decrement key
Note:
If none of the keys are pressed in operating mode then the system returns to the default display mode
3.5.
Reset to Factory Settings
In case of faulty configuration, the MP26transmitter can be reset to the default
manufacturer’s condition. The steps for resetting MP 26 is as given below
1.
Keep the keys increment and decrement pressed during power-on.
Then the display shows
wait for 1 sec then the display shows
2.
Then, press key increment to select YES.
3.
Confirm factory resetting with Enter and the copy procedure is started.
4.
Afterwards the device restarts.
3.5.1.
Working with MP26/10 tool
Connect the device to the tool using the cable supplied with the device. Then run the
MP26/10 software. A window prompt for selecting the project type appears.
Open an Exisiting project:
If this option is selected, it will open a window for selecting the
project.
Open last project:
If this option is selected, it will open the last project that was closed
earlier.