The underframe of the vessels must be strong enough to support the speciied loads,
horizontal (water level!) and lat.
-
Preferably, vessels should be installed on 3 weighing modules with integrated
stopper (see igure).
-
Parasitic and/or horizontal forces and torques exceeding the permissible limits are
disturbances which can generate measuring failures and, in the worst case, may
damage the load cell.
3.2 Foundation
Upper plates are not connected with each other.
Domed feet, non-ixated
Domed feet, welded
3 Recommendations for installation
Hygienic Weighing Module Contego®
Minebea Intec
EN-7