3.4.2
Jack-up
WARNING
The vessel can tip over.
Do not lift the vessel by more than 6 mm.
Lift the vessel as follows:
Use the load cell as follows:
1. Loosen the nut (5).
2. Turn the threaded bolt (4) upwards at the hex to lift the vessel.
3. Turn the threaded bolt (4) until the load cell is unloaded and can be removed
together with the upper load disc.
4. Insert the load disc into the load cell.
5. Insert the load cell into the mounting kit.
6. Push the load cell cable through the cable gland and block it; see Chapter
.
7. Load the load cell slowly and adjust the built-in lift-of protection as described in
Chapter
.
Hygienic Weighing Module Contego®
3 Recommendations for installation
EN-10
Minebea Intec