6
Combics CAPXS.U
Combics CAPXS.U
7
Installation
7
Installation
§
Connect the cable of the weighing platform to a suitable indicator
.
Note:
The cable gland along with the screw fasteners is already pre-assembled.
Use extreme care when attaching or detaching a cable.
§
Strip off the insulation at the cable end and attach the cable as follows:
– Route the cable through the cable gland.
– Properly tighten the screw fasteners of the cable gland
– Remove the sheathing from a section of the cable end (see illustration).
The shield (1) must have contact with the clamps (2).
– Expose approximately 15 cm (6 inches) of the wires (3) for installation.
§
Attach the cable to the weighing platform as follows:
– Expose approximately 5 cm (2 inches) of the wires for installation.
– Expose approximately 1 cm (1/2 inch) of the wires and attach ferrules to the wires.
– Securely attach the wires to the screw terminals
Wiring Diagram for the Indicator
No. Signal name
Meaning
Conductor coloring
1
BR_POS
Bridge supply voltage (+)
Blue
2
SENSE_POS
Sense (+)
Green
Bridge supply voltage
3
OUT_POS
Measuring voltage
(+)
White
4
OUT_NEG
Measuring voltage (-)
Red
5
SENSE_NEG
Sense (-)
Gray
Bridge supply voltage
6
BR_NEG
Bridge supply voltage (-)
Black or brown
§
Use a screwdriver to tighten the terminal screws.
77
(+)
Содержание Combics CAPXS U Series
Страница 19: ...Combics CAPXS U 19...
Страница 22: ...22 Combics CAPXS U...
Страница 23: ...Combics CAPXS U 23...