Combics CAPXS.U

 

Combics CAPXS.U

   

7

Conditions for Installation in Hazardous Locations

Before putting the equipment into operation, it is important to make sure that the cable

of the power supply is correctly connected to the power outlet (mains supply).

All equipment must be connected to the equipotential bonding conductor via grounding

cable (not included in delivery) connected to the grounding terminals on each device. 

The dimensions of the grounding cable are specified in national regulations for electrical

installations. Installation must be performed by a trained technician in accordance with

national regulations and generally acknowledged rules of engineering.

Use only cabling and extensions approved by 

Minebea Intec

, as these are made in accordance

with the restrictions on permissible cable lengths imposed by both the capacitance 

and inductivity values (see Control Drawing) and the requirements for electromagnetic

compatibility.

Before putting the weighing system into operation for the first time, make sure there 

is no hazard of explosion present at the place of installation. If there is any indication

that the equipment does not function properly (e.g., display remains blank, or no display

backlighting) due to damage during transportation, disconnect the equipment from

power and notify your nearest 

Minebea Intec

 Service Center.

The weighing platform specifications for Ui, Ii, Pi, Ta, temperature class, Ci and Li are 

listed in the enclosed Control Drawing. These parameters (Entity Concept) must be taken

into account when connecting an indicator to the platform. 

The explosion-protected weighing system must be installed in accordance with acknowl-

edged rules of engineering. These include national laws and regulations (e.g., National

Electrical Code).

Furthermore, national regulations for accident prevention and environmental protection

must be observed at all times.

Before the platform is operated in a hazardous location, it must be inspected either by 

a certified electrician or under the guidance and supervision of a certified electrician to

make sure that the weighing system complies with the applicable regulations. Determine

whether the installation must be registered with technical inspection authorities in your

country. Regular inspections must also be performed on the system during operation. 

The system should be inspected at intervals short enough to permit the prevention or

early detection of defects that arise as a result of normal wear and tear.

The longest permissible interval period is 3 years. Other conditions and standards that

regulate the installation and operation of the equipment and are applicable in your 

country must be met as well. When performing inspections, generally acknowledged rules

of engineering relevant to these conditions must also be applied.

If the terminal housing is opened by anyone other than persons authorized by 

Minebea Intec

,

or if the terminal is installed or operated incorrectly, this will result in forfeiture of the

approval for use in the stated hazardous location(s) and of all claims under the manufac-

turer’s warranty.

6

Содержание Combics CAPXS U Series

Страница 1: ...Operating Instructions Combics Minebea Intec Models CAPXS U Stainless Steel Weighing Platforms for Use in Hazardous Locations 98648 014 07 98648 014 07...

Страница 2: ...ety information in its entirety during installation and operation as well as while performing maintenance and repair work on the equipment It is important that all personnel using the Combics equipmen...

Страница 3: ...ebea Intec dealer or the Minebea Intec Service Center Any installation work that does not conform to the instructions in this man ual will result in forfeiture of all claims under the manufacturer s w...

Страница 4: ...ence when the equip ment is operated in a commercial envi ronment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction...

Страница 5: ...e leveling feet turn clockwise To lower the weighing platform retract the leveling feet turn counterclockwise After leveling the weighing platform retighten the locknuts securely as described below Lo...

Страница 6: ...ontrol Drawing These parameters Entity Concept must be taken into account when connecting an indicator to the platform The explosion protected weighing system must be installed in accordance with ackn...

Страница 7: ...ontact with the clamps 2 Expose approximately 15 cm 6 inches of the wires 3 for installation Attach the cable to the weighing platform as follows Expose approximately 5 cm 2 inches of the wires for in...

Страница 8: ...HING 3 OMBICS STAINLESS STEEL 0LATFORM IMENSIONS IN INCHES UMBER OF LOAD CELLS 3ELECTION OF THE WEIGHING CAPACITY IN LBS You should not place loads on the scale that exceed its maximum weighing capaci...

Страница 9: ...e use of steel wool sponges is not permitted Solvents are permitted for use only on stainless steel parts 10 10 Cleaning Unplug the scale from the AC power before cleaning To clean the weighing platfo...

Страница 10: ...4 30 24 YAR CWSS 30 30 30 YAR CWSS 30 36 36 YAR CWSS 36 48 48 YAR CWSS 48 60 48 YAR CWSS 48 60 48 YAR CWSS 60 60 60 YAR CWSS 60 11 Drive on ramp AISI 316 Ti stainless steel tread plate for platform si...

Страница 11: ...essory no DD 12 12 height 24 18 Adjustable to a max height of 25 31 YWT01CWP EE 18 18 height 24 18 Adjustable to a max height of 25 31 YWT02CWP Bench stainless steel for size Size in inch Accessory no...

Страница 12: ...he equipment described has been found to comply with the following Approval standards and other documents FM Class 3600 2018 FM Class 3610 2010 FM Class 3611 2018 FM Class 3810 2018 ANSI ISA 60079 0 2...

Страница 13: ...For Models CAPX IUX 4 and IFX 4 Class I II III Division 1 Groups A B C D E F and G T4 Ta 35739 003 07 A4 Entity Class I Division 2 Groups A B C D T4 Ta 50 C Class II III Division 2 Groups E F G T4 Ta...

Страница 14: ...IFX 4 Weighing Platforms Model YDI05 Z Interface Converter 13 Specific Conditions of Use 14 Test and Assessment Procedure and Conditions This Certificate has been issued in accordance with FM Approval...

Страница 15: ...as been found to comply with the following Approval standards and other documents CAN CSA C22 2 No 157 92 2006 CAN CSA C22 2 No 213 208 CAN CSA C22 2 No 142 2004 CAN CSA C22 2 No 60079 0 2007 CAN CSA...

Страница 16: ...II III Divisions 1 and 2 Groups A B C D E F and G as per control drawing 65710 800 07 A4 Entity Zone 20 Ex ia IIIC as per control drawing 65710 800 07 A4 Entity Zone 1 Ex ia IIC as per control drawin...

Страница 17: ...ntrinsically safe barrier and interface converter It is used to connect devices located in the non hazardous area with devices located in hazardous locations The signals that are exchanged between non...

Страница 18: ...rnpike Norwood MA 02062 USA T 1 1 781 762 4300 F 1 1 781 762 9375 E mail information fmapprovals com www fmapprovals com F 348 Mar 16 Page 4 of 4 10th March 2006 Original Issue 11th November 2018 Supp...

Страница 19: ...Combics CAPXS U 19...

Страница 20: ...s than 1 ohm In Canada The Apparatus must be connected to a suitable ground electrode per Canadian Electrical Code Part 1 The resistance of the ground pad must be less than 1 ohm 3 The Entity Concept...

Страница 21: ...cal Code NFPA 70 Article 504 or 505 The resistance of the ground pad must be less than 1 ohm In Canada The Apparatus must be connected to a suitable ground electrode per Canadian Electrical Code Part...

Страница 22: ...22 Combics CAPXS U...

Страница 23: ...Combics CAPXS U 23...

Страница 24: ...means without prior written permission from Minebea Intec All rights reserved The status of the information specifications and illustrations in this manual is indicated by the date given below Minebe...

Отзывы: