Minebea Intec Combics CAPXS U Series Скачать руководство пользователя страница 24

Minebea Intec Bovenden GmbH & Co. KG  
Leinetal 2 
37120 Bovenden, Germany

Phone   +49 (0) 551.309.83.0 
Fax  

+49 (0) 551.309.83.190

www.minebea-intec.com

Copyright by Minebea Intec, Bovenden, Germany. 
No part of this publication may be reprinted or 
translated in any form or by any means without prior 
written permission from Minebea Intec.  
All rights reserved.  
The status of the information, specifications and 
illustrations in this manual is indicated by the date given 
below. Minebea Intec reserves the right to make changes 
to the technology, features, specifications, and design 
of the equipment without notice.

Date: December 2018

Printed in Germany on paper that has been bleached 
without any use of chlorine. MS · KT
Publication No.: WCA6005-e18126

Содержание Combics CAPXS U Series

Страница 1: ...Operating Instructions Combics Minebea Intec Models CAPXS U Stainless Steel Weighing Platforms for Use in Hazardous Locations 98648 014 07 98648 014 07...

Страница 2: ...ety information in its entirety during installation and operation as well as while performing maintenance and repair work on the equipment It is important that all personnel using the Combics equipmen...

Страница 3: ...ebea Intec dealer or the Minebea Intec Service Center Any installation work that does not conform to the instructions in this man ual will result in forfeiture of all claims under the manufacturer s w...

Страница 4: ...ence when the equip ment is operated in a commercial envi ronment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction...

Страница 5: ...e leveling feet turn clockwise To lower the weighing platform retract the leveling feet turn counterclockwise After leveling the weighing platform retighten the locknuts securely as described below Lo...

Страница 6: ...ontrol Drawing These parameters Entity Concept must be taken into account when connecting an indicator to the platform The explosion protected weighing system must be installed in accordance with ackn...

Страница 7: ...ontact with the clamps 2 Expose approximately 15 cm 6 inches of the wires 3 for installation Attach the cable to the weighing platform as follows Expose approximately 5 cm 2 inches of the wires for in...

Страница 8: ...HING 3 OMBICS STAINLESS STEEL 0LATFORM IMENSIONS IN INCHES UMBER OF LOAD CELLS 3ELECTION OF THE WEIGHING CAPACITY IN LBS You should not place loads on the scale that exceed its maximum weighing capaci...

Страница 9: ...e use of steel wool sponges is not permitted Solvents are permitted for use only on stainless steel parts 10 10 Cleaning Unplug the scale from the AC power before cleaning To clean the weighing platfo...

Страница 10: ...4 30 24 YAR CWSS 30 30 30 YAR CWSS 30 36 36 YAR CWSS 36 48 48 YAR CWSS 48 60 48 YAR CWSS 48 60 48 YAR CWSS 60 60 60 YAR CWSS 60 11 Drive on ramp AISI 316 Ti stainless steel tread plate for platform si...

Страница 11: ...essory no DD 12 12 height 24 18 Adjustable to a max height of 25 31 YWT01CWP EE 18 18 height 24 18 Adjustable to a max height of 25 31 YWT02CWP Bench stainless steel for size Size in inch Accessory no...

Страница 12: ...he equipment described has been found to comply with the following Approval standards and other documents FM Class 3600 2018 FM Class 3610 2010 FM Class 3611 2018 FM Class 3810 2018 ANSI ISA 60079 0 2...

Страница 13: ...For Models CAPX IUX 4 and IFX 4 Class I II III Division 1 Groups A B C D E F and G T4 Ta 35739 003 07 A4 Entity Class I Division 2 Groups A B C D T4 Ta 50 C Class II III Division 2 Groups E F G T4 Ta...

Страница 14: ...IFX 4 Weighing Platforms Model YDI05 Z Interface Converter 13 Specific Conditions of Use 14 Test and Assessment Procedure and Conditions This Certificate has been issued in accordance with FM Approval...

Страница 15: ...as been found to comply with the following Approval standards and other documents CAN CSA C22 2 No 157 92 2006 CAN CSA C22 2 No 213 208 CAN CSA C22 2 No 142 2004 CAN CSA C22 2 No 60079 0 2007 CAN CSA...

Страница 16: ...II III Divisions 1 and 2 Groups A B C D E F and G as per control drawing 65710 800 07 A4 Entity Zone 20 Ex ia IIIC as per control drawing 65710 800 07 A4 Entity Zone 1 Ex ia IIC as per control drawin...

Страница 17: ...ntrinsically safe barrier and interface converter It is used to connect devices located in the non hazardous area with devices located in hazardous locations The signals that are exchanged between non...

Страница 18: ...rnpike Norwood MA 02062 USA T 1 1 781 762 4300 F 1 1 781 762 9375 E mail information fmapprovals com www fmapprovals com F 348 Mar 16 Page 4 of 4 10th March 2006 Original Issue 11th November 2018 Supp...

Страница 19: ...Combics CAPXS U 19...

Страница 20: ...s than 1 ohm In Canada The Apparatus must be connected to a suitable ground electrode per Canadian Electrical Code Part 1 The resistance of the ground pad must be less than 1 ohm 3 The Entity Concept...

Страница 21: ...cal Code NFPA 70 Article 504 or 505 The resistance of the ground pad must be less than 1 ohm In Canada The Apparatus must be connected to a suitable ground electrode per Canadian Electrical Code Part...

Страница 22: ...22 Combics CAPXS U...

Страница 23: ...Combics CAPXS U 23...

Страница 24: ...means without prior written permission from Minebea Intec All rights reserved The status of the information specifications and illustrations in this manual is indicated by the date given below Minebe...

Отзывы: