6 - 17
Test and Calibration
Nurse Call Performance Test
6.8
Nurse Call Performance Test
Required Tool: Multimeter
1.
Connect the nurse call cable to the analog output connector.
2.
If needed, press
to display the Normal Screen.
3.
Press
to display the
SYSTEM SETUP
dialog.
4.
Press
or
to highlight
MAINTENANCE
.
5.
Once
MAINTENANCE
is highlighted, press
to display the
MAINTENANCE
dialog shown in FIGURE 6-2.
6.
Press
or
to highlight
NURSE CALL
.
7.
Once
NURSE CALL
is highlighted, press
to display the
NURSE CALL SETUP
dialog shown in FIGURE 6-15.
FIGURE 6-15 NURSE CALL SETUP
dialog
8.
Press
or
to highlight the
SIGNAL DURATION
pull-down list.
9.
Once the
SIGNAL DURATION
pull-down list is highlighted, press
to select it.
10.
Press
or
to highlight the
PULSE
.
11.
Once the
PULSE
is highlighted, press
to select it.
12.
Press
or
to highlight the
SIGNAL TYPE
pull-down list.
13.
Once the
SIGNAL TYPE
pull-down list is highlighted, press
to select it.
14.
Press
or
to highlight the
NORMAL OPEN
.
15.
Once the
NORMAL OPEN
is highlighted, press
to select it.
16.
Press
or
to highlight either
HIGH
or
LOW
for
ALM LEV
.
17.
Once an
ALM LEV
is highlighted, press
to select it.
18.
Press
or
to highlight either
TECH
or
PHYS
for
ALM TYPE
.
19.
Once an
ALM TYPE
is highlighted, press
to select it.
20.
Trigger an alarm and measure the contact output with the multimeter. The output should
be square waves with an interval of 1s.