NetGuard Service Manual
0070-10-0685
2 - 15
Repair Information
Troubleshooting
6.
Select the ASCII Setup button and verify that the box for
Echo typed characters
locally
is checked. If not, check the box.
FIGURE 2-15
7.
Select
OK
. In the IP142 Properties window, select
OK
.
8.
At the OK prompt in the Hyperterminal window, enter the command
upgrade
. The
command
Transfer S-record file now, type ‘end’ when transfer complete.
OK
displays.
FIGURE 2-16
9.
Transmit the file by selecting
Transfer
from the menu bar at the top of the window, then
select
send text file
. Find and select the access point program file.
NOTE:
If the file is not a text file, go to Files of Type at the bottom
of the window and select “All files”.