Troubleshooting
Technical Messages and Solutions
BP10 Operator’s Manual
12 - 3
Clock Needs to Be Reset
Yes
The button cell does not have
sufficient charge.
Reset the system time.
NIBP Module Error
No
■
An error occurred to the
NIBP module.
■
There is a problem with
communications between
the module and the
telemetry.
Restart the device.
NIBP Cuff Loose
Yes
■
The NIBP cuff is not
properly connected.
■
There is a leak in the
airway.
1.
Check the patient’s
condition and verify
patient type.
2.
Replace with an appropri-
ate cuff and connect it
correctly.
NIBP Airway Error
Yes
The device cannot finish one
correct zeroing.
Restart the device.
If the error persists, contact your
service personnel.
NIBP Weak Signal
Yes
The patient’s pulse is weak or the
cuff is loose.
Check the patient condition and
change the cuff application site.
If the error persists, replace the
cuff.
NIBP Overrange
Yes
The measured NIBP value is not
within the measurement range.
Contact your service personnel.
NIBP Excessive Motion
Yes
Patient’s arm moves too much.
Check the patient’s condition
and reduce the patient motion.
NIBP Cuff Overpressure
Yes
The NIBP airway may be
occluded.
Check the airway and measure
again.
NIBP Cuff or Airway
Leak
Yes
The NIBP airway may leak air.
1.
Verify that the cuff is
properly connected.
2.
Verify that the airway
does not leak air.
NIBP Timeout
Yes
■
Time is out.
■
The measurement time is
over 120 seconds.
1.
Check the patient’s
condition and NIBP
connections.
2.
Replace the cuff.
NIBP Cuff and Patient
Mismatch
Yes
The cuff type applied mismatches
the patient category.
Verify the patient category and
replace the cuff.
Intervals Not Set
Yes
No interval in Sequence mode is
not set.
Set the intervals.
Message
Clearable
(Yes/No)?
Possible cause
Solution
Содержание BP10 NIBP Module
Страница 1: ...BP10 NIBP Module Operator s Manual...
Страница 2: ......
Страница 11: ...BP10 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 5...
Страница 16: ...Equipment Symbols Safety 1 6 BP10 Operator s Manual This page intentionally left blank...
Страница 38: ...Service Menu Configuring BP10 4 6 BP10 Operator s Manual This page intentionally left blank...
Страница 42: ...Discharging the Patient Patient Management 5 4 BP10 Operator s Manual This page intentionally left blank...
Страница 62: ...Disposing of the Batteries Battery 8 8 BP10 Operator s Manual This page intentionally left blank...
Страница 73: ...BP10 Operator s Manual 11 1 11 Accessories NIBP Accessories 11 2 Miscellaneous 11 3...
Страница 76: ...Miscellaneous Accessories 11 4 BP10 Operator s Manual This page intentionally left blank...
Страница 87: ...BP10 Operator s Manual B 1 B EMC...
Страница 93: ...BP10 Operator s Manual C 1 C FCC Compliance...
Страница 95: ...BP10 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3...
Страница 98: ...Abbreviations Symbols and Abbreviations D 4 BP10 Operator s Manual P power PR pulse rate...
Страница 99: ......
Страница 100: ...P N 046 011008 00 4 0...