Correcting the Measurement when Cuff is not at Heart Level
Monitoring NIBP
6 - 10
BP10 Operator’s Manual
3.
Countdown time
This time is available in automatic, continuous, sequence and ABPM modes.
4.
Mean pressure
5.
Pulse rate
6.
Measurement end time
6.9
Correcting the Measurement when Cuff is not at
Heart Level
Apply the cuff to a limb at the same level as the patient’s heart. If the cuff is not at the
heart level, do the following to the displayed value:
■
Add 0.75 mmHg (0.1 kPa) for each centimeter higher; or
■
Deduct 0.75 mmHg (0.1 kPa) for each centimeter lower.
6.10
Assisting Venous Puncture
You can use the NIBP cuff to cause sub-diastolic pressure to block the venous blood
vessel and therefore help the venous puncture.
To enable the venous puncture, follow this procedure:
1.
In the
Menu
screen, select
Venipuncture
.
2.
Verify the target pressure is appropriate. Change it if necessary.
3.
Select
Inflate
.
The display shows the real-time inflation pressure of the cuff, destination inflation
pressure, and the remaining time in venous puncture mode.
4.
Puncture vein and draw blood sample.
5.
When the puncture is complete, select
Deflate
, or press
key. The cuff
deflates automatically after a set time if you do not deflate it.
6.11
Reviewing NIBP Data
You can review the historic measurements at BP10 by selecting
History
in the main
menu.
BP10 can store and display up to 500 NIBP measurements. All the stored measurements
can resume if there is power interruption.
Содержание BP10 NIBP Module
Страница 1: ...BP10 NIBP Module Operator s Manual...
Страница 2: ......
Страница 11: ...BP10 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 5...
Страница 16: ...Equipment Symbols Safety 1 6 BP10 Operator s Manual This page intentionally left blank...
Страница 38: ...Service Menu Configuring BP10 4 6 BP10 Operator s Manual This page intentionally left blank...
Страница 42: ...Discharging the Patient Patient Management 5 4 BP10 Operator s Manual This page intentionally left blank...
Страница 62: ...Disposing of the Batteries Battery 8 8 BP10 Operator s Manual This page intentionally left blank...
Страница 73: ...BP10 Operator s Manual 11 1 11 Accessories NIBP Accessories 11 2 Miscellaneous 11 3...
Страница 76: ...Miscellaneous Accessories 11 4 BP10 Operator s Manual This page intentionally left blank...
Страница 87: ...BP10 Operator s Manual B 1 B EMC...
Страница 93: ...BP10 Operator s Manual C 1 C FCC Compliance...
Страница 95: ...BP10 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3...
Страница 98: ...Abbreviations Symbols and Abbreviations D 4 BP10 Operator s Manual P power PR pulse rate...
Страница 99: ......
Страница 100: ...P N 046 011008 00 4 0...