Configuring the NIBP Setup
Monitoring NIBP
6 - 6
BP10 Operator’s Manual
6.5.1
General NIBP Setup
6.5.2
Advanced NIBP Setup (Passcode Protected)
The following table lists the configurable options.
Options
Description
How to Enter
Settings*
Mode
Selects the NIBP measurement
mode.
Refer to
for additional
information.
Enter the main menu and
select
Mode
.
Manual
, Interval,
Sequence, ABPM
NIBP Color
Selects the NIBP parameter
color.
1. Enter the main menu.
2. Select
System
.
3. Select
NIBP Color
.
6 colors
The default color is white.
Initial Pressure
Selects the initial cuff inflation
pressure.
1. Enter the main menu.
2. Select
Patient Info
.
3. Select
Initial Pressure
.
Adult: 80 -280 mmHg.
Default: 160mmHg.
Pediatric: 80 - 210 mmHg
Default:140 mmHg
*
The factory default settings are in bold.
Options
Description
How to Enter
Settings*
NIBP Timeout
Configures the retention time of the
measurement value each time it is
obtained.
If
OFF
is selected, the measurement
value retains till the next measurement
starts.
1. Enter the main menu.
2. Select
System
→
Maintenance
. A passcode
screen displays.
3. Enter the correct
passcodes.
4. Select
NIBP Settings
.
5 min, 10 min,
15 min, 30 min,
45 min, 60 min,
OFF
NIBP Endtone
Configures it to issue a reminder tone at
the completion of a NIBP measurement.
ON,
OFF
Measure on
Clock
Configures it to automatically take an
NIBP measurement on the fixed time
based on the interval.
For example, in a situation that
Measure
on Clock
is set to
ON
, and
Interval
is
20
min
, if the first measurement is taken at
14:03, the following measurements will
be taken at 14:20, 14:40, 15:00 and so on.
ON
, OFF
Units
Configures the NIBP measurement unit.
mmHg,
kPa
*
The factory default settings are in bold.
Содержание BP10 NIBP Module
Страница 1: ...BP10 NIBP Module Operator s Manual...
Страница 2: ......
Страница 11: ...BP10 Operator s Manual 1 1 1 Safety Safety Information 1 2 Equipment Symbols 1 5...
Страница 16: ...Equipment Symbols Safety 1 6 BP10 Operator s Manual This page intentionally left blank...
Страница 38: ...Service Menu Configuring BP10 4 6 BP10 Operator s Manual This page intentionally left blank...
Страница 42: ...Discharging the Patient Patient Management 5 4 BP10 Operator s Manual This page intentionally left blank...
Страница 62: ...Disposing of the Batteries Battery 8 8 BP10 Operator s Manual This page intentionally left blank...
Страница 73: ...BP10 Operator s Manual 11 1 11 Accessories NIBP Accessories 11 2 Miscellaneous 11 3...
Страница 76: ...Miscellaneous Accessories 11 4 BP10 Operator s Manual This page intentionally left blank...
Страница 87: ...BP10 Operator s Manual B 1 B EMC...
Страница 93: ...BP10 Operator s Manual C 1 C FCC Compliance...
Страница 95: ...BP10 Operator s Manual D 1 D Symbols and Abbreviations Units D 2 Symbols D 3 Abbreviations D 3...
Страница 98: ...Abbreviations Symbols and Abbreviations D 4 BP10 Operator s Manual P power PR pulse rate...
Страница 99: ......
Страница 100: ...P N 046 011008 00 4 0...