background image

P/N: H-046-007060-00 Rev 3.0

9

BeneVision TD60

Quick Reference Guide

ECG 

Select ECG Leadwire

The TD60 allows the clinician to set the ECG Leadwire Type being used. To 
configure the ECG Leadwire Type:

• 

On the main screen, tap the HR digital area or ECG waveform area to

          enter the 

ECG

 menu.

OR

  1.  Press   to enter the main menu.

  2. Tap 

Parameter Setup

.

  3. Tap 

ECG

.

  4. Select 

Cable Type

.

   5.  Select

• 

Auto

: the device automatically sets the leadwire type according to the 

              leads connected.

• 

3 Lead

: to select 3-leadwire set

              If the leadwire type is set to 3-lead, the 

Smart Lead

 option becomes

              

Monitored Lead

. The clinician can select the preferred lead from the

              

Monitored Lead

.

• 

5 Lead

: to select 5-leadwire set

              All ECG leads display on the main screen.

  6. Press

 

 to return to the main screen.

Lead Placement

The TD60 Lead Placement menu allows the clinician to check the ECG Lead 
Status and troubleshoot Lead Fault messages. To enter the Lead Placement 
menu:

• 

Tap the lead off message on the main screen.

OR

• 

Press 

 to enter the main menu, and then tap 

Lead Placement

.

 

Содержание BeneVision TD60

Страница 1: ...Quick Reference Guide BeneVision TD60 Display Telepack...

Страница 2: ...ended as a replacement to the Operating Instructions Prior to operating equipment the user must be familiar with the Operating Instructions Manual contents This document is a guideline only to be used...

Страница 3: ...nt Category 6 Standby 7 Enter Standby 7 Resume Monitoring 8 Printing 8 Nurse Call 8 Activate a Nurse Call 8 Acknowledge the Nurse Call 8 ECG 9 Select ECG Leadwire 9 Lead Placement 9 Enable Pacer 10 Re...

Страница 4: ...tion 16 Flip the Landscape Display 16 Adjust Display Order 17 Screen Lock View Only Mode 17 Enable the Screen Lock 17 Unlock the Screen in Locked Mode 18 Unlock the Screen in View Only Mode 18 Adjust...

Страница 5: ...turn the display on Press and hold this key for two seconds to display the power off confirmation menu 2 Nurse Call key Nurse call key activates nurse call message at the Central Monitoring System CMS...

Страница 6: ...viewing patient information and adjusting patient settings 5 Patient alarm light indicator 6 ECG lead connector 7 SpO2 cap Covers SpO2 connector when SpO2 is not in use 8 SpO2 connector Connects the...

Страница 7: ...P N H 046 007060 00 Rev 3 0 3 BeneVision TD60 Quick Reference Guide GENERAL NAVIGATION Swipe for More Data Sw ipe S w i p e Sw ipe S w i p e...

Страница 8: ...nce Guide View More Data The TD60 provides the following shortcuts to access additional information as shown in the following figure Tap to enter the Patient Info menu Tap to to view the MPAN info Tap...

Страница 9: ...P N H 046 007060 00 Rev 3 0 5 BeneVision TD60 Quick Reference Guide Quick Key Screen Menu Bar Deployment Note The quick keys are configurable This image is just representative S w i p e u p Press...

Страница 10: ...ecting that a new patient is on the TD60 Note Discharging the patient on the TD60 discharges the patient from the CMS Select the Discharge Patient menu 1 Press to enter the main menu 2 Tap Discharge P...

Страница 11: ...default preset settings but does not clear patient information or data Note When the device is connected to the CMS the patient category at the CMS is updated if the patient category is changed on th...

Страница 12: ...to the CMS To print a strip Select Print from the quick key area or Command section in the main menu NURSE CALL Activate a Nurse Call Press to activate a nurse call request to the CMS The alarm light...

Страница 13: ...the leadwire type according to the leads connected 3 Lead to select 3 leadwire set If the leadwire type is set to 3 lead the Smart Lead option becomes Monitored Lead The clinician can select the pref...

Страница 14: ...lead The color of the flashing circle is based on the alarm level Enable Pacer The TD60 allows the clinician to enable pacemaker detection The symbol displays in the waveform area on the CMS when the...

Страница 15: ...will trigger the No Chest Movement alarm To set NCM Alarm Delay On the main screen tap the Resp digital area or Resp waveform area to enter the Resp menu OR 1 Press to enter the main menu 2 Tap Param...

Страница 16: ...ick Reference Guide SpO2 OPTIONAL The TD60 allows continuous pulse oximetry Connect the SpO2 Module Connect the SpO2 module to the TD60 as shown in the following figure Connecting the Masimo SpO2 modu...

Страница 17: ...to the SpO2 module and the SpO2 module to the TD60 2 Select an appropriate sensor according to the patient size and weight 3 Prepare the SpO2 site in accordance with hospital policy 3 Apply the senso...

Страница 18: ...f devices If not repeat Steps 1 to 4 5 In the list of devices select your desired device and tap Connect Once the device is connected successfully the status will be changed from Disconnect to Connect...

Страница 19: ...he main menu 2 Tap Wireles Modules 3 Disable MPAN Unpair a TD60 with a BP10 via the BP10 1 Press the MPAN key on the BP10 2 Select Yes when the prompt message Are you sure you want to close MPAN and u...

Страница 20: ...pear before the alarm message The symbol appears on the top of the main screen For the technical alarm The alarm sound will be silenced The alarm light will continue to indicate the alarm The alarm wi...

Страница 21: ...return to the main screen SCREEN LOCK VIEW ONLY MODE The TD60 provides the ability to lock the touch screen display to restrict user access options for View Only and Locked are available View Only al...

Страница 22: ...in View Only Mode 1 If the display is off press or to activate the display 2 Press to display the Screen Locked menu 3 Enter the passcode to unlock the screen Once the passcode is entered the screen i...

Страница 23: ...r color configuration on the TD60 and the CMS are independent CLEAN DISINFECT THE TD60 Clean the TD60 Clean your equipment on a regular basis Before cleaning consult your hospital s regulations for cl...

Страница 24: ...OPA Solution Ortho Phthalaldehyde 0 55 Metrex CaviCide1 Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0 28 Isopropanol 17 2 Virex II 256 Didecyl dimethyl ammonium chloride 8 704 n Al...

Страница 25: ...rox Healthcare Hydrogen Peroxide Cleaner Disinfectant Wipes Hydrogen Peroxide 1 4 Diversey Oxivir TB Wipes Hydrogen Peroxide 0 5 Metrex CaviWipes Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium...

Страница 26: ...are available for use For equipment without the symbol only the cleaning and disinfecting agents marked with are available for use To clean the equipment 1 Shut down the device 2 Clean the display sc...

Страница 27: ...nfect reusable ECG leadwires and SpO2 sensors refer to the instructions for use delivered with the accessory STERILIZATION Sterilization is not recommended for this equipment related products accessor...

Страница 28: ...Mindray DS USA Inc 800 MacArthur Boulevard Mahwah New Jersey 07430 USA Tel 1 800 288 2121 1 201 995 8000 www mindray com 0 4 6 0 0 7 0 6 0 0 0 3 0...

Отзывы: