P/N: H-046-007060-00 Rev 3.0
14
BeneVision TD60
Quick Reference Guide
NIBP
(OPTIONAL)
The TD60 allows non-invasive blood pressure (NIBP) monitored by a BP10 NIBP
module.
Pair a TD60 with a BP10
1. Press the MPAN key on the right panel of the BP10. The“Pairing...”
message will display in the message area of the BP10.
2. Press to enter the main menu.
3. Tap
Wireless Modules
.
4. Enable
MPAN
.
The TD60 starts searching the BP10 devices in the vicinity and
displays the devices that can be connected under
Devices
. Make sure
that the device you wish to connect appears in the list of devices. If not,
repeat Steps 1 to 4.
5. In the list of devices, select your desired device and tap
Connect
. Once
the device is connected successfully, the status will be changed from
Disconnect
to
Connected
.
6. Tap anywhere in the NIBP digital area on the main screen to access the
NIBP
menu.
7. Tap
Start
to start NIBP measurement. Verify that the NIBP measurement
results displayed at the BP10 are consistent with that displayed on the
main screen of the TD60 and are the data for the same patient.
Pair a TD60 with a New BP10
1. Unpair the TD60 with the BP10 already connected by following steps in
Unpair a TD60 with a BP10 via the TD60
or
Unpair a TD60 with a
BP10 via the BP10 .
2. Pair the TD60 with your desired BP10 by following steps in
Pairing a
TD60 with a BP10.