E - 1
E
Voice Prompts
The following table lists voice prompts that may occur during a rescue.
Condition
Voice Prompt
Description
Open the lid
Powered on. Stay calm. Follow the
instructions.
The lid is opened.
Device error. Recommended to replace
the Device. Stay calm. Follow the
instructions.
The equipment malfunctions, use
one standby equipment or start CPR
immediately.
After turning on the
equipment
Adult mode
The Adult/Child mode switch is
pressed to Adult, or the electrode
pads connected to the equipment
are detected for the adult patient.
Child mode. if the patient is an Adult,
adjust the Adult/Child mode switch to
Adult mode.
The Adult/Child mode switch is
pressed to Child.
Child mode
The Adult/Child mode switch is
pressed to Child, or the electrode
pads connected to the equipment
are detected for the children.
Place the electrode
pads
Remove clothing from patient's chest.
Apply pads as shown on Pads.
Prompts to remove the patient’s
clothing and place the electrode
pads.
Remove clothing from patient's chest.
Plug in pads connector.
Remove pads package from lid of
AED.Tear open package. Apply pads as
shown on Pads.
Apply pads as shown on Pads.
Apply pads as shown on Pads.
Abnormal Pads connection.
Pads connection failure, start CPR
immediately.
The equipment
analyzes the patient’s
heart rhythm.
Do not touch the patient. Analyzing heart
rhythm.
Repeats until analysis of the
patient’s heart rhythm is completed.
This prompt will be interrupted if
the equipment is ready to shock.
No shock advised.
Notifies non-shockable rhythm has
been detected.
Motion detected. Do not touch or move
the patient.
The equipment detects ECG noise
artifacts, stop moving or touching
the patient.
Noise detected. Make sure pads are firmly
attached.
The equipment detects ECG noise
artifacts, better pads contact on the
patient’s skin is required.
Pads off. Analysis interrupted.
Pads connection failure, the
equipment automatically stops the
heart rhythm analysis. Reconnect
the electrode pads.
Содержание BeneHeart C Series
Страница 2: ......
Страница 10: ...4 This page intentionally left blank ...
Страница 14: ...1 4 This page intentionally left blank ...
Страница 22: ...3 4 This page intentionally left blank ...
Страница 34: ...6 4 This page intentionally left blank ...
Страница 40: ...8 4 This page intentionally left blank ...
Страница 42: ...9 2 9 2 Miscellaneous Description Model PN Disposable battery LM34S002A 022 000411 00 ...
Страница 50: ...A 8 This page intentionally left blank ...
Страница 64: ...E 4 This page intentionally left blank ...
Страница 68: ...F 4 This page intentionally left blank ...
Страница 70: ...G 2 This page intentionally left blank ...
Страница 72: ...H 2 This page intentionally left blank ...
Страница 74: ......
Страница 75: ...I 1 I Declaration of Conformity ...
Страница 76: ...I 2 ...
Страница 77: ......
Страница 78: ...P N 046 012619 00 2 0 ...