A - 7
A.11
ECG Specifications (for Equipment Configured with the Screen)
A.12
Electrode Pads Specifications
ECG inputs
Multifunction electrode pads
Gain
Auto
Sweep speed
25 mm/s, error no more than
±
5%
Common mode rejection
>90 dB
Recovery time
<2.5 s (after defibrillation)
Electrode Pads
MR60
MR61
MR63
MR62
Electrode shape
Oval
Cable length
1.2 m preconnectable
Total area
148 cm
2
75 cm
2
75 cm
2
126 cm
2
Adhesive Area
145 cm
2
74 cm
2
74 cm
2
123 cm
2
Maximum Number of
Defibrillation Shocks
Up to 50 shocks (360J monophasic and biphasic)
Shelf life (with sealed package)
36 months
60 months
Storage Condition
0ºC to 50ºC
15ºC to 35ºC
The shelf life assumes a storage
temperature of 25ºC. Storage
temperature above 25ºC will reduce
the shelf life.
Содержание BeneHeart C Series
Страница 2: ......
Страница 10: ...4 This page intentionally left blank ...
Страница 14: ...1 4 This page intentionally left blank ...
Страница 22: ...3 4 This page intentionally left blank ...
Страница 34: ...6 4 This page intentionally left blank ...
Страница 40: ...8 4 This page intentionally left blank ...
Страница 42: ...9 2 9 2 Miscellaneous Description Model PN Disposable battery LM34S002A 022 000411 00 ...
Страница 50: ...A 8 This page intentionally left blank ...
Страница 64: ...E 4 This page intentionally left blank ...
Страница 68: ...F 4 This page intentionally left blank ...
Страница 70: ...G 2 This page intentionally left blank ...
Страница 72: ...H 2 This page intentionally left blank ...
Страница 74: ......
Страница 75: ...I 1 I Declaration of Conformity ...
Страница 76: ...I 2 ...
Страница 77: ......
Страница 78: ...P N 046 012619 00 2 0 ...