2.2
Limitations of use
Limitations of use
•
Under federal law (only applicable to the USA) this product may only be
sold by or on the order of a physician or licensed practitioner.
•
The product may only be used under the constant supervision of or on the
instructions of a physician or other authorized medical professional.
The product is NOT intended for:
•
critical patient monitoring
•
use in life support systems
The product is NOT to be:
•
used near MRI equipment
•
exposed to ionizing radiation
•
used on patients undergoing electro surgery
•
used in oxygen rich environments (concentration > 25 % at 1 atm)
The product is NOT:
•
suitable for use in an inflammable mixture of anesthetics or agents and air,
oxygen or nitrous oxide
•
defibrillator proof
•
suitable for sterilization
Do not use, store or transport the equipment outside the specified environmental
conditions, this may damage the product.
Do not store or use in environments with Magnetic Resonance Imaging (MRI)
equipment, or equipment capable of emitting diagnostic levels of ionizing radiation.
Apart from the above, there are no contra-indications. There are no known side effects from the use
of this product.
2.3
Safety measures and warnings
Warnings
•
Do not combine the use of the product with any other electronic equipment,
except those specified in this manual. Doing so may impair the product’s
emissions and immunity regarding EMC.
•
The product can only be used with the accessories designated by the
manufacturer. The use of other accessories may impair the product's
emissions and immunity regarding EMC.
•
The accessories supplied with the device can only be used with Mind Media
approved devices.
•
Sensors with their own power are not to be connected to any of the inputs.
•
Transmission quality decreases when there are other radio devices in the
neighborhood. The wireless transmission may be interfered with by other
equipment.
•
The product should not be used adjacent to or stacked with other
equipment. If this is required, then it should be observed if normal
operation of the product in that configuration is confirmed.