![MindMedia NeXus-4 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/mindmedia/nexus-4/nexus-4_user-manual_1800742016.webp)
16
7
Electromagnetic guidance
Portable and mobile RF communications equipment can affect the system. The system needs special
precautions regarding EMC and must be installed and put into service according to the EMC
information outlined below.
Guidance and manufacturer’s declaration - electromagnetic emissions
The NeXus-4 is intended for use in the electromagnetic environment specified below. The customer or the user of the
NeXus-4 should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The NeXus-4 uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions CISPR 11
Class B
The NeXus-4 is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions IEC 61000-3-2
Not applicable
Battery powered
equipment
Voltage fluctuations/flicker emissions
61000-3-3
Not applicable
Battery powered
equipment
Guidance and manufacturer’s declaration - electromagnetic immunity
The NeXus-4 is intended for use in the electromagnetic environment specified below. The customer or the user of the
NeXus-4 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
± 8 kV contact
± 8 kV air
Floors should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
Not applicable
Battery powered
equipment
Not applicable
cabling shorter than
3m
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Not applicable
Battery powered
equipment
Not applicable
Battery powered
equipment