Place Minicap electrodes on to the red EXG sensors TP and place a Minicap electrode on to
the EXG ground TP. Slightly twist the connector in to EXG TP connection in order to make good
connection.
Place black Minicap electrodes on to the black electrodes of the EXG sensor TP which is
connected to input A&B of the NeXus-10 and input 25&26 of the NeXus-32.
Place the black TP connection (reference) electrodes on to an ear clip.
Determine electrode positions according to the 10-20 system by using measuring tape and
marking positions with a skin marker.
Use the clip of the EXG TP cable to attach the EXG TP sensor to prevent pulling of the cables.
Place the Minicap or Microcap on the head.