Temperature Indicator Calibration 

Calibration  of  the  temperature  indicator  can  be  done  in  a  number  of  ways;  comparison  to 
another  calibrated  instrument,  using  decade  box  simulation,  or  in  a  controlled  environment 
such as an oil bath.  The TI408 series indicator incorporates a calibration menu to adjust the 
zero offset for each channel. 
 
Controlled Environment / Comparison Method  
The  most  accurate  method  of  re-calibrating  the  indicator  /  sensor  system  is  to  place  the 
sensor(s) in  a known  environment such  as  an oil-bath or  a calibration chamber and base all 
measurements  off  of  a  known  standard.    Calibration  using  this  method  will  account  for  any 
lead-length  or  contact  resistance,  as  well  as  any  offset  characteristics  of  the  sensor.    The 
environment should have an accurate sensor or instrument (standard) with an accuracy of less 
than or equal to 0.1°C from which all measurements will be based.  To compare the calibration 
of the indicator to another calibrated instrument or sensor, both sensors should be within 1

” of 

each  other.    Set  the  temperature  of  the  environment  to  the  mid-point  of  the  desired 
measurement range.  For example, if all measurements fall within the range of 120°C to 135°C, 
set  the  environment  to  127.5°C.  Wait  for  the  indicator  and  standard  to  stabilize  before 
proceeding.    Once  the  system  is  stable,  follow  the  calibration  procedures  below  for  each 
channel to calibrate the indicator.   
 
Decade Box 
Connect a  resistance  decade box with a resolution of at  least  .01 ohms to the desired input 
(CH1-

CH8) of the indicator.  If unsure or concerned about the decade box’s accuracy, measure 

the desired  set-point using an accurate  4-wire  ohmmeter.  Set  the decade box  resistance to 
simulate the mid-point of the desired measurement range.  For example, if all measurements 
fall within the range of 120°C to 135°C, simulate the resistance for 127.5°C (1488.92

W

 for a PF 

type sensor).  Follow the instructions below to calibrate the indicator for each channel.  
 
 

 
 
 
 
 
 
 
 
 
 
 

Содержание TI408

Страница 1: ...Tel 763 571 3121 Fax 763 571 0927 www minco com TI408 8 Channel Handheld Temperature Indicator Operating Instructions...

Страница 2: ...erance Specifications Display 2 line 3 digit LCD Liquid Crystal Display Indicating Range 40 0 C to 160 0 C Sensing Elements 1000W platinum RTD TCR 0385 W W C 1000 W 0 C per channel Sensor Input Indica...

Страница 3: ...Minco Tel 763 571 3121 Fax 763 571 0927 www minco com 3 Temperature Indicator Functionality...

Страница 4: ...1 C from which all measurements will be based To compare the calibration of the indicator to another calibrated instrument or sensor both sensors should be within 1 of each other Set the temperature o...

Страница 5: ...Minco Tel 763 571 3121 Fax 763 571 0927 www minco com 5 Calibration Procedures...

Страница 6: ...Minco Tel 763 571 3121 Fax 763 571 0927 www minco com 6...

Страница 7: ...3121 Fax 763 571 0927 www minco com 7 How to Order TI408 Model Number TI408 8 Channel Handheld Temperature Indicator PF Element Type PF 1000W Platinum RTD TCR 0385 W W C 1000W at 0 C Default TI408PF S...

Страница 8: ...exclusive remedy available to the buyer in the event of a breach by the seller Items that show evidence of mishandling or misapplication may be returned by the seller at the customer s expense Further...

Страница 9: ...Minco Tel 763 571 3121 Fax 763 571 0927 www minco com 9 Dimensions...

Страница 10: ...Desk Usine et Service Minneapolis MN Tel 1 763 571 3123 Commercial Z I 55432 Fax 1 763 571 0942 09310 Aston France USA custserv minco com Tel 33 5 61 03 24 01 Tel 1 763 571 3121 www minco com Fax 33 5...

Отзывы: