MANUFACTURER DIRECTIVES AND DECLARATION – ELECTROMAGNETIC IMMUNITIES

Our devices are designed to be used in an ELECTROMAGNETIC ENVIRONMENT as specified below. The user must 
ensure that the units can operate in that environment. 

NOTE:

 Home health care settings have higher immunity requirements than professional medical care facilities. Therefore, 

the interference immunity requirements for professional medical care centers are included here.

STANDARD BASIC EMC 

IMMUNITIES TEST

TEST LEVEL

IEC 60601

COMPLIANCE

ELECTROMAGNETIC ENVIRONMENT

Electrostatic Discharge (ESD)
IEC (EN) 61000-4-2

±8kV

Contact

±15kV

Air

±8kV

Contact

±15kV

Air

Floors must be made from wood, 
concrete or ceramic tiles.
If the floors are covered in synthetic 
material, the relative humidity must be at 
least 30%.

Electrical Fast Transients 
(bursts)
IEC (EN) 61000-4-4

±2kV for power lines

±1kV between 

phases

±2kV for power lines

±1kV between 

phases

The quality of the electrical supply 
network must be equivalent to that of a 
commercial or hospital setting.

Mains surge 
IEC (EN) 61000-4-5

±2kV between 

phases and earth

±1kV  for input/

output/earth lines

±2kV between 

phases and earth

±1kV  for input/

output/earth lines

The quality of the electrical supply 
network must be equivalent to that of a 
commercial or hospital setting.

Voltage Drops  
IEC (EN) 61000-4-11

0% UT (100% drop 

in UT); 0.5 cycles

40% UT (60% drop 

in UT); 10 cycles

0% UT (100% drop 

in UT); 5s

0% UT (100% drop 

in UT); 0.5 cycles

40% UT (60% drop 

in UT); 10 cycles

0% UT (100% drop 

in UT); 5s

The quality of the electrical supply 
network must be equivalent to that 
of a commercial or hospital setting. 
If the device user requires continued 
use during power cuts, we recommend 
powering the device with a different 
power source or battery.

Magnetic Field at Industrial 
Frequency
IEC (EN) 61000-4-8

30 A/m

30 A/m

Magnetic fields at industrial frequency 
must be those of a typical hospital or 
commercial setting.

Radiated Radiofrequency

IEC (EN) 61000-4-3

10 V/m 

80MHz - 6GHz

10 V/m 

80MHz - 6GHz

The distance between portable or 
mobile radiofrequency communication 
equipment and the device, including 
cables, must not be less than the 
recommended separation distance, 
which is calculated using the applicable 
equation for the transmitter frequency.

Recommended separation distance:

d

 = 1,2√

P

; < 80MHz

d

 = 1,2√

P

 ; de 80MHz - 800MHz

d

 = 2,3√

P

 ; de 800MHz - 2,7GHz

P

 is the maximum rated power output 

of the transmitter in watts (W) and 

d

 is 

the recommended separation distance in 
meters (m).

Conducted Radiofrequency

IEC (EN) 61000-4-6

10V

150kHz - 100MHz

10V

150kHz - 100MHz

NOTE 1: 

UT is the AC network voltage before applying the test level.

NOTE 2:

 The upper frequency range is applied at 80MHz and 800MHz.

NOTE 3:

 These directives may not be applicable in all situations. Electromagnetic propagation is affected by the absorption 

and reflection of structures, objects and people.

Содержание M183500

Страница 1: ...ENG USER MANUAL MS EXAMINATION LAMP MS LED MS LED PLUS MS FLEX MS FLEX PLUS MS CEILING MS CEILING PLUS F3P30 Rev 01 20 06 2022...

Страница 2: ...U REGULATION 2017 745 on medical devices amended and integrated This manual s content may be modified in part or in whole by MISMAL without prior notice in order to make changes and improvements MIMSA...

Страница 3: ...MS model 16 5 Safety functions 18 5 1 Voltage drop 18 5 2 Power cuts 18 5 3 Electrical malfunction 18 6 Cleaning Disinfection 18 6 1 General safety instructions 18 6 2 Cleaning 18 6 3 Disinfection 19...

Страница 4: ...ted humidity values of between 10 and 75 for transportation and storage WARNING Not doing so could result in serious or even fatal injuries CAUTION Not doing so could result in minor to moderate injur...

Страница 5: ...source unit 2 The lighting unit does not include a fail safe power source or an emergency battery 3 In the event of a power cut the lighting unit will shut off completely 4 Short black outs are possi...

Страница 6: ...m knocks Collision with other objects could result in the failure of the device and or damage to the cover and the support arm system causing parts to fall off The lighting unit does not include a fai...

Страница 7: ...ore starting to install the device check that the content is in good condition and that it has not been damaged or degraded during transportation Claims will only be accepted if the vendor or the carr...

Страница 8: ...aint Hold the pole 1 firmly and turn the screw 7 with a wrench or a scanner until the whole set is tight Luminaries fixation system 8 Allen number 2 2 Technical characteristics Length of pole 75 cm To...

Страница 9: ...attached Once the clamp is fixed and secure position the lamp through the upper hole 5 Install the accessory without the lighting unit installed on it It is the customer s responsibility to ensure tha...

Страница 10: ...ws until the assembly the guide separator and the screw are firmly attached 2 Once the support is fixed and secure position the lamp through the upper hole 3 Install the accessory without the lighting...

Страница 11: ...firmly attached 3 Once the support is fixed and secure position the lamp through the upper hole 4 Install the accessory without the lighting unit installed on it It is the customer s responsibility to...

Страница 12: ...the other end of the extension arm 4 Turn the extension arm s side screw in order to secure the luminaire 5 3 5 EXTENSION ARM REF MODEL DESCRIPTION B0101S0 EXTENSION ARM Extension arm including Wall...

Страница 13: ...ION CEILING SUPPORT Ceiling bracket included in the MS CEILING and MS CEILING PLUS models Not compatible with MS LED MS LED PLUS MS FLEX and MS FLEX PLUS Install the accessory without the lighting uni...

Страница 14: ...and insert screws 4 Attachment and anchorage 1 Position the support on the ceiling 2 Mark the 5 support holes to install the attachment 5 Electrical connection 1 Insert the cable into the bottom of t...

Страница 15: ...ion whilst verifying the main function and the control system 4 2 OPERATION OF THE LIGHTING UNIT HEAD Lighting unit with simple and ergonomic operation for intuitive handling Do not use the unit if th...

Страница 16: ...X MS FLEX PLUS 44 cm 141 cm 92 cm 49 cm 95 5 cm 16 44 cm 160 cm 92 cm 58 cm 15 cm 63 cm 130 cm 105 cm 66 cm 1 Minimum length 2 Maximum height 3 Maximum length 4 Minimun height 1 Minimum length 2 Maxim...

Страница 17: ...ENG MS CEILING MS CEILING PLUS 100 cm 200 cm 120 cm 180 cm 85 cm 61 cm 25 cm 68 cm 45 cm 112 cm 128 cm 126 cm 17 1 Minimum length 2 Maximum height 3 Maximum length 4 Minimun height...

Страница 18: ...them 4 Apply the cleaner by wetting a cloth never by applying it directly to the device 6 2 CLEANING SAFETY Check the general safety instructions RECOMMENDED CLEANING 1 Use a mild soap solution as a...

Страница 19: ...Disinfectants may contain harmful substances that could injure the skin or eyes or damage respiratory organs if inhaled 7 MAINTENANCE Medical devices must undergo regular cycles of maintenance and rev...

Страница 20: ...ct an authorized recycling company NOTE Do not dispose of the product with the usual domestic waste Take all the disinfection and or sterilization measures before disposing of the device so as to not...

Страница 21: ...ING MS CEILING PLUS Box Dimensions 265 x 800 x 180 mm 210 x 1000 x 160 90 mm 260 x 250 x 1045 mm Gross Weight 3 64 kg 3 62 kg 2 70 kg 2 68 kg 6 18 kg 6 16 kg Lighting unit Net Weight 2 28 kg 2 26 kg 1...

Страница 22: ...s have higher immunity requirements than professional medical care facilities Therefore the interference immunity requirements for professional medical care centers are included here EMISSIONS TESTS S...

Страница 23: ...0 UT 100 drop in UT 0 5 cycles 40 UT 60 drop in UT 10 cycles 0 UT 100 drop in UT 5s 0 UT 100 drop in UT 0 5 cycles 40 UT 60 drop in UT 10 cycles 0 UT 100 drop in UT 5s The quality of the electrical su...

Страница 24: ...ossible to resolve the incident remotely in which case it will respond to the Buyer with instructions on how to proceed or conversely if the Product must be sent to MIMSAL to be examined and if necess...

Страница 25: ...emarks that are identical or similar to those used by MIMSAL or that could cause confusion for customers regarding the identity or character of MIMSAL or of the Products It is expressly forbidden to u...

Страница 26: ...A12 2015 Part 1 Medical equipments UNE EN 60601 1 2 2015 Part 2 Medical equipments UNE EN 60601 2 41 2010 A11 2012 A1 2015 B0101800 WALL B SUPPORT B010190 AH TABLE CLAMP 82026 RAIL PLUS BRACKET 09651...

Страница 27: ...Catalanes 130 136 Pl 9 08038 Barcelona Spain TEL 34 93 479 26 00 www dnv es assurance MANAGEMENT SYSTEM CERTIFICATE Certificate no 283445 2019 AQ IBE ENAC Initial certification date 20 April 2016 Val...

Страница 28: ...EL 39 68 99 905 www dnv it MANAGEMENT SYSTEM CERTIFICATE Certificate no 283446 2019 AQ IBE ACCREDIA Initial certification date 27 December 2017 Valid 09 February 2022 08 February 2025 This is to certi...

Страница 29: ...ENG MIMSAL TRADE S L C Mollet 17 Pol gono Industrial Palou Nord 08401 Granollers Barcelona SPAIN Tel 34 930 139 860 mimsal mimsal com www mimsal com F3P30 Rev 01 20 06 2022...

Отзывы: