PRELIMINARY EXAMINATION

Each meter is supplied complete with:

y

SE600 preamplified non refillable double junction

combination pH electrode with built in EC/TDS probe and 

temperature sensor; 1 m (3.3’) cable

y

Calibration solutions, 20 mL each

 

▶ pH 7.01, 1413 µS/cm, 1382 ppm (MW801)

 

▶ pH 7.01, 1413 µS/cm, 1500 ppm (MW802)

y

Instruction manual

y

1 x 9 V battery 

OPERATION

Slide off the battery compartment cover located on the back 

of the meter, and install the battery into the battery clip 

connector while observing polarity.

Connect the probe to the meter securely by aligning the pins 

with the socket and pushing the plug in.

DIN Connector

Always remove the electrode protective cap before use.

Make sure that the meter has been calibrated before taking 

any measurements (see calibration procedure).

Turn the instrument on by pressing the ON/OFF key.

ON/OFF

pH MEASUREMENTS

If the electrode has been left dry, soak the tip in MA9015 

Storage Solution for a few hours (or overnight) to reactivate.

Submerge the tip of the electrode into the sample to be 

measured, select the pH mode and allow the reading to 

stabilize before taking measurements.

EC/TDS MEASUREMENTS

Immerse the tip of the electrode into the sample to be tested.

Select the measurement range (EC or TDS) and wait for the 

temperature sensor to reach thermal equilibrium with the 

samples before taking measurements.

EC/TDS measurements have a high thermal dependency 

and readings taken with this instrument are temperature 

compensated.

AFTER MEASUREMENTS the instrument should be switched 

off and the probe cleaned and stored with a few drops of 

storage solution in the protective cap.

CALIBRATION PROCEDURE

Remove the protective cap from the electrode.

Switch the meter on.

Select the range pressing the appropriate key for pH, EC or TDS.

For best calibration results we suggest calibration to be 

done at room temperature (around 25 

°

C/77 

°

F).

pH Calibration
A. PREPARATION
Single point manual calibration. Choose buffer.

If you are going to measure samples near pH 7 use pH 7.01 

buffer solution (M10007B) for calibration; use pH 4.01 buffer 

solution (M10004B) for acidic measurements, or pH 10.01 

(M10010B) for alkaline measurements.

M10007B

B. PROCEDURE
Rinse the tip of the electrode and immerse it in the 

calibration solution. Wait a couple of minutes for the reading 

to stabilize.

Adjust the calibration knob until the LCD shows the pH value 

at the value printed on the used buffer standard.

pH Calibration

Calibration is now complete and the meter is ready for use.
EC/TDS Calibration
A. PREPARATION
Single point EC calibration.

Choose appropriate standard:

y

MW801:   use  1382 ppm (M10032B) for TDS or

use 1413 µS/cm (M10031B) for EC calibration.

y

MW802:   use  1500 ppm (M10442B) for TDS or 

use 1.413 mS/cm (M10031B) for EC calibration.

Calibration

Solution

B.

 

PROCEDURE

Immerse the tip of the electrode in the standard solution. 

Wait for the reading to stabilize, and thermal equilibrium be 

reached. 

Turn the EC/TDS calibration knob until the display shows 

the EC or TDS reading at 25 °C

,

 

value printed

 on the 

calibration bottle’s or sachet’s label you are using.

EC/TDS Calibration

Calibration is now complete and the meter is ready for use.

The instrument should be re-calibrated at least once a 

month, or when the probe or battery is replaced.
Note: The conversion between EC and TDS is made by a 

built-in circuit. If calibration is made in EC range, TDS range 

is automatically calibrated (or vice versa).

BATTERY REPLACEMENT

The meters are powered by a 9  V battery that is in a 

compartment located in the rear of the instrument.

When the battery becomes weak, the instrument 

automatically switches off.

Replacement must only take place in a non-hazardous area 

using an alkaline 9 V battery. 

Slide off the battery compartment cover and replace the old 

battery with a new one. Make sure that the battery contacts 

are fully engaged in the connector. Seat the battery in is 

compartment and replace the cover.