MW170 Bench Meter
Messages Displayed During Measurement
y
If the EC measurement exceeds the specified limits or the temperature exceeds the
– 20 to 120 °C (– 4.0 to 248.0 °F) range, “OUT OF SPEC” message is displayed on the
third LCD line.
y
If the EC measurement exceeds the user-selected range, the “OVER RANGE” message
is displayed on the third LCD line.
y
“NO CAL” message indicates that the probe needs to be calibrated or that the
previous calibration has been deleted.
y
If the probe is not connected, “NO PROBE” message is displayed on the third LCD line.
Messages Displayed During Interval Logging
y
If EC temperature exceeds the specified limits, “OUT OF SPEC” message is displayed
alternatively with the Log specific messages.
y
If the probe sensor is disconnected or damaged, logging stops and "NO PROBE"
message is displayed on the third LCD line. The log file will indicate "Log end - Probe
disconnected".