Rinse the electrode tip with distilled water, immerse it (4 cm / 1½") into the sample and
stir gently for a few seconds. For a faster response and to avoid cross contamination of
the samples, rinse the electrode tip with the solution to be tested before taking any
measurements.
STO RAG E
PRO CEDURE
STO RAG E
PRO CEDURE
STO RAG E
PRO CEDURE
STO RAG E
PRO CEDURE
STO RAG E
PRO CEDURE
To minimize clogging and ensure a quick response time, the glass bulb and the junction
should always be kept moist.
When not in use, store it with a few drops of M
M
M
M
MA 9015
A 9015
A 9015
A 9015
A 9015 storage solution (not included with
Bench Meters) in the protective cap.
PERIO DIC
MAIN TENAN CE
PERIO DIC
MAIN TENAN CE
PERIO DIC
MAIN TENAN CE
PERIO DIC
MAIN TENAN CE
PERIO DIC
MAIN TENAN CE
Inspect electrode and cable. The cable used for the connection to the instrument must
be intact and there must be no points of broken insulation on the cable or cracks on the
electrode stem or bulb. If any scratches or cracks are present, replace the electrode.
Rinse off any salt deposits with water.
Connectors must be perfectly clean and dry.
For refillable electrodes:
Refill the electrode with fresh electrolyte (see the electrode’s specifications to select the
correct refilling solution). Allow the electrode to stand upright for 1 hour. Follow the
Storage Procedure above.
CLEANING
PRO CEDURE
CLEANING
PRO CEDURE
CLEANING
PRO CEDURE
CLEANING
PRO CEDURE
CLEANING
PRO CEDURE
• General:
Soak in MA 9016
MA 9016
MA 9016
MA 9016
MA 9016 General Cleaning Solution for approx. ½ hour.
(not included with Bench Meters)
IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thoroughly
with distilled water and soak it in M
M
M
M
MA
A
A
A
A 9015
9015
9015
9015
9015 storage solution for at least 1 hour before taking
measurements.
NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER