CALIBRATION:

•  Remove the protective cap from the electrode. 

•  Immerse the electrode tip into a new sachet of pH 7 calibration 

solution and allow the reading to stabilize.

•  Adjust the pH7 calibration trimmer (on the front) to display “7.0 

pH”. 

•  Rinse the electrode with tap water and dry it with paper tissue.

•  Open a new sachet of pH 4 calibration solution, immerse the 

electrode tip into the sachet and allow the reading to stabilize. 

•  Adjust the pH4/10 calibration trimmer (on the front) to display 

“4.0 pH”.

The calibration is now complete and the meter is ready to take 

measurements. 

It is suggested to recalibrate the meter at least once a month, 

after a prolonged stocking time and after pH-electrode 

 

replacement.

INSTALLATION OF THE 

MP810 1,5 L/H DOSING PUMP

Connect the rigid tube to the pump (aspiration – arrow upwards). 

Fix the white filter on the other end of the tube, where the dis

-

solved pH reducing acid will be drawn in – to prevent small ob-

jects entering the pump.

Connect the soft tube to the pump (impulsion – arrow down-

wards). The injection fitting will be placed at the end of this tube 

only to inject into a piping. (This fitting has a thread that links it 

to the piping.)

INSTALLATION OF CONTROLLER:

Connect the supplied 240VAC/12VDC power adapter to the  

meter and to the main line.

Connect the connection base (black and white outlet) to the 

main line.

Connect the pH-electrode to the BNC socket on the bottom of 

the meter.

Always remove the electrode protective cap before taking any 

measurement. Make sure that the meter has been calibrated 

before taking any measurements

Immerse the tip (2.5 cm) of the pH-electrode into the sample. 

(for example it can float with the help of a piece of polystyrol).

Turn the instrument on by pressing the ON/OFF button.

The nature of the setpoint can be selected by setting the switch 

to the desired position (ABOVE or BELOW).

Allow the reading to stabilize and the meter will start continuous 

monitoring.

A blinking alarm will indicate when the measured pH value is 

higher or lower than selected setpoint, depending on user  

selection. When the alarm is activated, it also activates the  

connection base and the auxiliary equipment.

The setpoint can be selected by adjusting the central front knob 

to the desired value. The selectable range is from 5.5 to 9.5 pH. 