background image

8

  MA881 Digital Invert Sugar Refractometer

4. PRINCIPLE OF OPERATION

The Invert sugar determination is made by measuring the refractive 

index of a solution. Refractive Index is an optical characteristic of a 

substance and the number of dissolved particles in it. Refractive 

Index is defined as the ratio of the speed of light in empty space 

to the speed of light in the substance. A result of this property is 

that light will “bend”, or change direction, when it travels through 

a substance of different refractive index. This is called refraction.
When passing from a material with a higher to lower refractive 

index, there is a critical angle at which an incoming beam of 

light can no longer refract, but will instead be reflected off the 

interface. The critical angle can be used to easily calculate the 

refractive index according to the equation:

sin (

θ

critical

) = n

2

 / n

1

Where n

2

 is the refractive index of the lower-density medium; n

1

 is 

the refractive index of the higher-density medium.

i

r

n1

n2

 = angle of incidence

 = angle of refraction

n1 , n2 -

  refractive index

Refraction

Critical Angle

Total Internal Reflection

r

i

i

critical

i

r

n2

n1

n2

i

i

critical

i

critical

 < 

 > 

 = 

n1

θ
θ

θ

θ

θ

θ

θ

θ

θ

θ

θ

θ

θ

In the 

MA881

, light from an LED passes through a prism in contact 

with the sample. An image sensor determines the critical angle 

at which the light is no longer refracted through the sample. The 

MA881

 automatically applies temperature compensation to the 

measurement and converts the refractive index of the sample to 

Invert sugar concentration in units of percent (by weight).

Prism

Lens

Linear Image Sensor

Sample

Shadow

Led

Sample Well

Содержание MA881

Страница 1: ...INSTRUCTION MANUAL SATISFACTION GUARANTEED MA881 Digital Invert Sugar Refractometer...

Страница 2: ...is prohibited without the written consent of the copyright owner Milwaukee Instruments Inc Rocky Mount NC 27804 USA THANK YOU for choosing Milwaukee Instruments This instruction manual will provide yo...

Страница 3: ...CRIPTION 6 3 SPECIFICATIONS 7 4 PRINCIPLE OF OPERATION 8 5 MEASUREMENT GUIDELINES 9 6 CALIBRATION PROCEDURE 10 7 MEASUREMENT PROCEDURE 12 8 CHANGING TEMPERATURE UNIT 14 9 ERROR MESSAGES 15 10 BATTERY...

Страница 4: ...OGRESS TAG C SETUP FACTORY CALIBRATION TAG D CAL CALIBRATION TAG E PRIMARY DISPLAY DISPLAYS MEASUREMENT AND ERROR MESSAGES F AUTOMATIC TEMPERATURE COMPENSATION BLINKS WHEN TEMPERATURE EXCEEDS 10 40 C...

Страница 5: ...Refractometer 0 to 85 Mass A E F G B C D A LIQUID CRYSTAL DISPLAY LCD B READ KEY USER MEASUREMENT C ZERO KEY USER CALIBRATION D ON OFF E PRIMARY DISPLAY F SECONDARY DISPLAY G STAINLESS STEEL SAMPLE W...

Страница 6: ...measurement technique and temperature compensation employ methodology recommended in the ICUMSA Methods Book Internationally recognized body for Sugar Analysis Temperature in C or F is displayed simul...

Страница 7: ...y 1 5 seconds Minimum Sample Volume 100 L cover prism totally Sample Cell SS ring and flint glass prism Temperature Compensation Automatic between 10 and 40 C 50 to 104 F Case Material ABS Enclosure R...

Страница 8: ...ill instead be reflected off the interface The critical angle can be used to easily calculate the refractive index according to the equation sin critical n2 n1 Where n2 is the refractive index of the...

Страница 9: ...measure seawater solutions Do not expose instrument or prism to solvents that will damage it This includes most organic solvents and extremely hot or cold solutions y Particulate matter in a sample ma...

Страница 10: ...n release Two instrument test screens will be displayed briefly an all segment screen followed by the percentage of remaining battery life When LCD displays dashes the instrument is ready 2 Using plas...

Страница 11: ...GES section Note The 0 0 screen will remain until a sample is measured or the instrument is turned off 4 Gently absorb the ZERO water standard with a soft tissue Use care not to scratch the prism surf...

Страница 12: ...le onto the prism surface Fill the well completely Note If the temperature of the sample differs significantly from the temperature of the instrument wait approximately 1 minute to allow thermal equil...

Страница 13: ...nsation is disabled if the temperature exceeds the 10 40 C 50 104 F range 4 Remove sample from the sample well by absorbing on a soft tissue 5 Using plastic pipettes rinse prism and sample well with d...

Страница 14: ...andholdtheON OFFkeycontinuouslyforapproximately 8 seconds The LCD will display the all segment screen followed by a screen with the model number on the primary display and the version number on the se...

Страница 15: ...bration used to zero instrument Use deionized or distilled water Press Zero HI Top display CAL segment ON Wrong calibration used to zero instrument Use deionized or distilled water Press Zero t LO Top...

Страница 16: ...f battery life is remaining Temperature values are blinking 0 0 C or 80 0 C Temperature measurement out of sampling range 0 0 to 80 0 C ATC segment blinking Outside temperature compensation range 0 to...

Страница 17: ...strument OFF by pressing the ON OFF key y Turn instrument upside down and remove the battery cover by turning it counterclockwise y Extract the battery from its location y Replace with fresh 9V batter...

Страница 18: ...hisproduct makesureitisentirelysuitableforyourspecific application and for the environment in which it is used Any modification introduced by the user to the supplied equipment may compromise the mete...

Страница 19: ...MilwaukeeInstrumentsreservestherighttomakeimprovementsindesign construction and appearance of its products without advance notice...

Страница 20: ...C H 6726 Szeged HUNGARY tel 36 62 428 050 fax 36 62 428 051 www milwaukeeinst com e mail sales milwaukeeinst com Milwaukee Instruments Inc 2950 Business Park Drive Rocky Mount NC 27804 USA tel 1 252 4...

Отзывы: