6
Menu Button>Wi-Fi Settings>View Device IP
When Wi-Fi is set up and you have logged in, you will be able to
view the Device IP and MAC Addresses.
Menu Button>Wi-Fi Settings>Upload Interval
This screen will show upload interval, that is, how often the
DL122 sends new information to the portal. Five minutes is the
minimum upload interval.
Menu Button>Wi-Fi Settings>Portal Sync
Menu Button>Wi-Fi Settings>Portal Service
IMPORTANT: Do not open the Portal Service screen unless
you are instructed to do so by Milwaukee Instruments. If you
accidentally open this screen, press MENU to back out. Do NOT
enter any data! Doing so will permanently disconnect you
from the portal.
Setting up the Incremental
Dosing Control (IDC)
The Incremental Dosing Control offers precision on/off switching
of pumps, lights, and other devices. (It’s a little sophisticated for
your coffee maker, but it’ll work.) It can be programmed for 1 to
7 days of the week with Start/End times each day and up to 999
timed-duration sessions per day. The controller function and IDC
are completely independent of each other.
Menu Button>IDC Setup>Days to Dose
1) Press Menu button, scroll down to **IDC Setup**, and press
SELECT.
2) The **IDC Setup** screen allows you to accurately schedule
precision dosing and includes a flow calibration feature.
Begin by placing the cursor beside “Days to Dose,” press SELECT.
3) Use the left/right arrows (<>) to move the blinking cursor
across the days of the week. On each day, use the up/down
arrows to toggle the day ON or OFF. The presence of an “X”
indicates that power will be supplied to a dosing device (or any
other device) on that day.
4) Once you have completed selection of days, press SELECT
and this information will be SAVED!
Menu Button>IDC Setup>Sessions/Day
The DL122 will space sessions evenly over the available time
(Start Time to End Time) on each day.
1) Press Menu button, scroll down to **IDC Setup**, and press
SELECT.
2) Place cursor beside Sessions/Day and press SELECT.
3) Use the left/right arrows (<>) to move the blinking cursor
across the three digits. In each place, use the up/down arrows
to select a number. Once you have entered the appropriate
number of sessions per day (up to 999), press SELECT and this
information will be SAVED!
4) Based on Start/End times and Session Time you have
programmed, if you attempt to enter a number of sessions that
will not fit into the time that is available, the message on your
screen will read: “Please enter a valid number of sessions per
day.” You will need to enter a lower number of Sessions Per Day
or make adjustments in Start/End times and/or Session Time.
Menu Button>IDC Setup>Session Time
Session Time is the length of time (duration) that your device
will be turned on. All session times must be the same length.
1) Press Menu button, scroll down to **IDC Setup**, and press
SELECT.
2) Place cursor beside Session Time and press SELECT.
3) Use the left/right arrows to move the blinking cursor across
the time, shown in Hours:Minutes:Seconds (00:00:00). Use the
up/down arrows to select numbers. Once you have entered the
appropriate duration of EACH dosing session, press SELECT
and the SAVED! screen will appear. All session times must be
the same.
4) Based on Start/End times and Sessions/Day you have
programmed, if you attempt to enter a session time that will
not fit into the time that is available, the message on your
screen with read: “Please enter a valid session time.” You will
need to enter a shorter session time or make adjustments in
Start/End times and/or Sessions/Day
Menu Button>IDC Setup>Start Time
1) Press Menu button, scroll down to **IDC Setup**, and press
SELECT.
2) Place cursor beside “Start Time” and press SELECT.
3) Use the left/right arrows to move the blinking cursor across
the start time and up/down arrows to select appropriate
numbers. When complete, press SELECT and the SAVED! screen
will appear. Note that the DL122 uses military (24-hour) time.
A start time of 2:30PM would be entered as 14:30. Default Start
Time is 00:00.
4) If you attempt to enter Start/End times that do not allow
sufficient time for Sessions/Day and Session Time you have
programmed, the message on your screen will read: “Please
enter valid Start and End Times.” You will need to add more time
between Start and End Times, if possible, or make adjustments
in Session Time and/or Sessions/Day.
Menu Button>IDC Setup>End Time
1) Press Menu button, scroll down to **IDC Setup**, and press
SELECT.
2) Place cursor beside “End Time” and press SELECT.
3) Use the left/right arrows to move the blinking cursor across
the end time and up/down arrows to select appropriate
numbers. When complete, press SELECT and the Saved! screen
will appear. Default End Time is 23:59.
The unit will not accept an End Time that precedes a Start Time,
and Start/End Times must fall within a single day.
4) If you attempt to enter Start/End times that do not allow
** IDC Setup **
> Days to Dose
Sessions/Day
Session Time
Start Time
End Time
Flow Calibration
Exit