Maintenance
33
Chapter 4
M
AINTENANCE
4-1
S
CHEDULED
M
AINTENANCE
T
ABLE
See the Maintenance Table below for the typical maintenance that needs to be performed on your Elix System. The catalogue
numbers and other ordering information are found in Section 6-1. Detailed information about the various consumable items can
be found in Section 2-5.
Item or action
Maintenance Needed
When?
How to?
A10 TOC Monitor
Calibration.
Once per year.
Contact Millipore SAS.
A10 UV Lamp
Replacement.
When prompted to by a LCD
Message.
See Section 4-9.
Clean A10 TOC Monitor
Cleaning.
When a QGard is changed.
See Section 4-8.
Clean RO Cartridges
Cleaning.
As necessary.
See Section 4-6.
Other Pretreatment
See Pretreatment Device User
Manual for this information.
See Pretreatment Device User
Manual for this information.
See Pretreatment Device User
Manual for this information.
PrePak L1 Pre-System
Pretreatment Pack
Replacement.
Change when Progard TL1 or
TL2 Pack is changed.
See information that came
with the Pack.
Progard TL1 or TL2 System
Pretreatment Pack
Replacement.
When prompted to by a LCD
Message.
See Section 4-2.
QGard TL Polisher Pack
Replacement.
When prompted to by a LCD
Message.
See Section 4-7.
RO Pump Pressure
Adjustment.
Water temp. < 25
°
C, operate
at 10 bar. (up to 12 bar with
cold feedwater).
Water temp.
≥
25
°
C, operate
at 7 bar.
Contact Millipore SAS.
See Section 4-4.
Sanitise RO Cartridges
Sanitisation.
Automatically when the
Progard TL1 or TL2 Pack is
changed or as needed.
See Section 4-5.
UV Lamp
Replacement.
When prompted to by a LCD
Message.
See Section 4-3.