Millipore Sigma Pellicon 3 Cassettes Скачать руководство пользователя страница 13

Pellicon

®

 3 Cassettes Installation and User Guide 

13

NOTE 

If the system is complex and has other associated manifolds, ensure that all 

wetted surfaces in the manifolds are exposed to the solution. All valves that have 

been exposed to process fluids should also be exposed. It is good practice to 

cycle (partially open and the partially close) valves at least twice over the course 

of the cycle to ensure that all wetted internal surfaces of the valve body are 

exposed to the solution.

8.  Filters may be stored in holders with the feed, retentate and filtrate isolation valves 

closed and enough compression on the cassettes to ensure that the fluid cannot 

escape or evaporate. 

9.  The filters may also be removed from the system and placed in sealed bags 

or containers.  Submerge in storage solution in appropriate container. Filters 

should be kept at 2-8 ºC. Process holders with hydraulics can properly hold and 

maintain storage solution.  Process holders are also designed with a store mode 

compression setting that releases to about 60% of the process compression and 

ensures that the units will not leak externally.

NOTE 

The storage agent used must comply with applicable local regulations. 

Recommended Storage Solutions for Pellicon

® 

 3 Cassette Filters

Membrane

Storage Agent

Concentration

pH

Recommended 

Time Period

Ultracel

®

NaOH*

0.05 N

12.7

1 year

NaOH

0.1 N 

13

6 months

Biomax

®

Acetic/Phosphoric Acid 0.16M Acetic/0.12M 

Phosphoric Acid

2 - 3

1 year

NaOH

0.1 N 

13

1 year

*Use milder conditions if effectiveness of agent is still observed. Better membrane life 

is expected with lower concentration of caustic and/or at 20-25 °C for regenerated 

cellulose membranes with a nominal molecular weight limit of 3 or 5 kDa.
These recommended storage agents will ensure membranes remain wet and will 

prevent microbiological growth without damage to the filter. Upon reinstallation, the 

filters should be flushed, cleaned and sanitized prior to use. 
Filters should be stored in liquid-tight bags or in containers at 2 - 8 °C and should 

not be frozen. Fresh solutions should be prepared after the recommended time period 

to prevent possible microbial growth or 

contamination. 
Ultracel

® 

 membranes can be stored in 

0.1 N NaOH only for the recommended 

time period. Longer exposure may result in 

membrane damage.

Содержание Pellicon 3 Cassettes

Страница 1: ...ale Holder Installation 4 Hydraulic Process Scale Holder Installation 5 Maintenance 7 Flushing and Cleaning 7 Flushing 8 Cleaning 8 Sanitization 9 Depyrogenation 10 Integrity Testing 10 Measurement of...

Страница 2: ...TE Pellicon 3 88 cm2 and 0 11 m2 Cassettes must be installed in a Pellicon Cassette Holder Catalog Number XX42PMINI Holder Installation 1 Loosen the nuts on the tie rods of the holder and remove the n...

Страница 3: ...wipe 4 Slide the manifold adapter plate onto the holder against the holder s manifold plate with the solid light grey side labeled Manifold Side towards the manifold and the wings resting on the tie...

Страница 4: ...ary Tri Clover connection Pellicon Cassette hex nut washer stainless steel bushing handle PERMEATE RETENTATE FEED Pellicon10_2_06_2 eps Gasket Plate manifold adapter plate nut PERMEATE R E T E N T A T...

Страница 5: ...wrench and 15 16 in deep hex socket turn each nut 1 8 turn in the sequence shown in Figure 4 When the wrench clicks this indicates that the set force has been reached DO NOT OVER TORQUE Evenness of to...

Страница 6: ...d tight 4 Turn on the air supply to the Hydraulic Power Unit A minimum of 80 psig air pressure should be available Check the air pressure on the pressure gauge PI 1 on Hydraulic Power Unit front panel...

Страница 7: ...ut replacement of nuts and washers should be considered after every ten installations to maintain high sealing force in the holder 7 When new nuts do not spin freely refurbish the threaded rods with a...

Страница 8: ...se the tank drain valve V1 and the tank valve V2 2 Fill the tank with the flushing solution 3 Fully open the retentate valve V4 and permeate valve V5 if present and direct both the retentate and filtr...

Страница 9: ...iofilms Biopolymer Proteins and Polyphenolic Membrane Cleaning Agent Concentration Temperature C pH Time min Ultracel NaOH 0 1 N 0 5 N 25 50 13 13 7 30 60 Biomax NaOH 0 1 N 0 5 N 25 50 13 13 7 30 60 N...

Страница 10: ...c Acid 0 1M 30 50 1 30 NaOH 0 1N 25 50 13 30 Ultracel NaOH 0 1N 25 50 13 30 Use milder conditions if effectiveness of agent is still observed Better membrane life is expected with lower concentration...

Страница 11: ...l the tank with filtered deionized water water for injection or reverse osmosis permeate The flush water must be extremely pure to avoid fouling the membranes or introducing other contaminants into th...

Страница 12: ...0 40 0 734 75 2 24 1 023 46 4 8 1 551 102 2 39 0 748 73 4 23 1 047 44 6 7 1 598 100 4 38 0 762 71 6 22 1 072 42 8 6 1 648 98 6 37 0 777 69 8 21 1 098 41 0 5 1 699 Based on Water Fluidity Relative to...

Страница 13: ...ess compression and ensures that the units will not leak externally NOTE The storage agent used must comply with applicable local regulations Recommended Storage Solutions for Pellicon 3 Cassette Filt...

Страница 14: ...completely flushed Flush system and cassette System and cassette not completely cleaned Clean system and cassette Symptom Possible Cause Remedy High NWP Damage to membrane or cassette Integrity test...

Страница 15: ...xed below fluid air interface Vortex in feed container Add baffle to break vortex or reduce mixing speed Retentate splashing Fix retentate line below fluid in feed air interface container Pump cavitat...

Страница 16: ...atory matters to the best of our knowledge and ability but without obligation or liability Existing laws and regulations are to be observed in all cases by our customers This also applies in respect t...

Отзывы: