Содержание Pellicon 3 Cassettes

Страница 1: ...ale Holder Installation 4 Hydraulic Process Scale Holder Installation 5 Maintenance 7 Flushing and Cleaning 7 Flushing 8 Cleaning 8 Sanitization 9 Depyrogenation 10 Integrity Testing 10 Measurement of...

Страница 2: ...TE Pellicon 3 88 cm2 and 0 11 m2 Cassettes must be installed in a Pellicon Cassette Holder Catalog Number XX42PMINI Holder Installation 1 Loosen the nuts on the tie rods of the holder and remove the n...

Страница 3: ...wipe 4 Slide the manifold adapter plate onto the holder against the holder s manifold plate with the solid light grey side labeled Manifold Side towards the manifold and the wings resting on the tie...

Страница 4: ...ary Tri Clover connection Pellicon Cassette hex nut washer stainless steel bushing handle PERMEATE RETENTATE FEED Pellicon10_2_06_2 eps Gasket Plate manifold adapter plate nut PERMEATE R E T E N T A T...

Страница 5: ...wrench and 15 16 in deep hex socket turn each nut 1 8 turn in the sequence shown in Figure 4 When the wrench clicks this indicates that the set force has been reached DO NOT OVER TORQUE Evenness of to...

Страница 6: ...d tight 4 Turn on the air supply to the Hydraulic Power Unit A minimum of 80 psig air pressure should be available Check the air pressure on the pressure gauge PI 1 on Hydraulic Power Unit front panel...

Страница 7: ...ut replacement of nuts and washers should be considered after every ten installations to maintain high sealing force in the holder 7 When new nuts do not spin freely refurbish the threaded rods with a...

Страница 8: ...se the tank drain valve V1 and the tank valve V2 2 Fill the tank with the flushing solution 3 Fully open the retentate valve V4 and permeate valve V5 if present and direct both the retentate and filtr...

Страница 9: ...iofilms Biopolymer Proteins and Polyphenolic Membrane Cleaning Agent Concentration Temperature C pH Time min Ultracel NaOH 0 1 N 0 5 N 25 50 13 13 7 30 60 Biomax NaOH 0 1 N 0 5 N 25 50 13 13 7 30 60 N...

Страница 10: ...c Acid 0 1M 30 50 1 30 NaOH 0 1N 25 50 13 30 Ultracel NaOH 0 1N 25 50 13 30 Use milder conditions if effectiveness of agent is still observed Better membrane life is expected with lower concentration...

Страница 11: ...l the tank with filtered deionized water water for injection or reverse osmosis permeate The flush water must be extremely pure to avoid fouling the membranes or introducing other contaminants into th...

Страница 12: ...0 40 0 734 75 2 24 1 023 46 4 8 1 551 102 2 39 0 748 73 4 23 1 047 44 6 7 1 598 100 4 38 0 762 71 6 22 1 072 42 8 6 1 648 98 6 37 0 777 69 8 21 1 098 41 0 5 1 699 Based on Water Fluidity Relative to...

Страница 13: ...ess compression and ensures that the units will not leak externally NOTE The storage agent used must comply with applicable local regulations Recommended Storage Solutions for Pellicon 3 Cassette Filt...

Страница 14: ...completely flushed Flush system and cassette System and cassette not completely cleaned Clean system and cassette Symptom Possible Cause Remedy High NWP Damage to membrane or cassette Integrity test...

Страница 15: ...xed below fluid air interface Vortex in feed container Add baffle to break vortex or reduce mixing speed Retentate splashing Fix retentate line below fluid in feed air interface container Pump cavitat...

Страница 16: ...atory matters to the best of our knowledge and ability but without obligation or liability Existing laws and regulations are to be observed in all cases by our customers This also applies in respect t...