48
Safety information
Never open the system, even when the power switch has been switched off.
HIGH VOLTAGE INSIDE!
Symbol
What it means
This UV RADIATION sticker is used to refer to a position on the
Water System Cabinet or inside of it where exposure to UV light
is possible
This DANGER sticker is used to refer to a position on the Water
System Cabinet or inside of it that could be hazardous.
This ELECTRICAL GROUND sticker is used to refer to a position on the
Water System Cabinet or inside where an electrical ground connection is
made.
This ELECTRICAL DANGER sticker is used to refer to a location on or
inside the Milli-Q system where an electrical danger could exist.
This CAUTION sticker is used to refer to a surface that can be hot.
Disconnect and switch power off to allow surface to cool before servicing.
Hazard
The Milli-Q system must be connected to a source of electrical power that
is earth grounded.
Attention
Before the system is serviced, unplug the electrical power cord. The
Milli-Q system must be powered OFF before plugging in or removing any
components on the electronic board(s).
The Milli-Q IQ 7003/7005/7010/7015 has been tested by an independent and accredited company
for compliance with EU directives related to safety and electromagnetic compatibility. The
declaration of conformity is available upon request. The system has been manufactured using
components and practices recommended by UL and has been cULus marked. The registration
and CB certificates can be verified at
.
In case of wall mounted installation:
For the system Milli-Q IQ 7003/7005/7010/7015 : Use a wall capable of supporting a maximum
load of 120 kg. Use 6mm diameter stud and fasteners capable to support a maximum load of
60 Kg each.
For the 25L tank: Use a wall capable of supporting a maximum load of 130 kg. Use 6mm diam
-
eter stud and fasteners capable to support a maximum load of 45 Kg each.
For the 50L tank: Use a wall capable of supporting a maximum load of 240 kg. Use 6mm diam
-
eter stud and fasteners capable to support a maximum load of 80 Kg each.
For the 100L tank: Use a wall capable of supporting a maximum load of 500 kg. Use 6mm diam
-
eter stud and fasteners capable to support a maximum load of 180 Kg each.