Self Help Guide
34
Low flowrate
•
Ensure the POD-Pak is not air-locked. This can happen in the weeks that follow a
replacement of IPAK cartridges. Dispense water and open the POD-Pak air vent to see if
there is any trapped air. Close the vent once the air has been released.
•
One possible reason for a decrease in Milli-Q Water flowrate is a clogged POD-Pak. The
POD-Pak should be replaced when it appears to be clogged.
Tank is not re-filled of pure water
•
Open the drop down menu swiping down on the touchscreen to check the status of the
pure production section.
•
Exit the maintenance or rinsing mode in case it is activated.
•
Solve any alarms that could be stoping the production process.