Страница 3: ...5 7010 7015 is intended to produce pure type 2 and ultrapure type 1 water from a tap water source primarily for use in research and quality control in a variety of laboratories worldwide The product i...

Страница 4: ...5 consists of different units 1 POD Point of Dispense with screen interface 6 Electrical connections 2 POD dispenser 7 Hydraulic connections 3 POD base 8 Storage tank 4 Dispensing wheel 9 Milli Q IQ E...

Страница 5: ...schematic Distribution flowschematic 1 2 3 5 6 9 4 7 8 10 12 11 13 1 Tap Feed Water 18 Pure water Resistivity Cell 2 Inlet Solenoid Valve 19 Pure water 3 Ways Valve 3 System Feed Pressure Sensor 20 e...

Страница 6: ...aminants The ech2 o ASM lamp further protects the integrity of the stored water with regular exposure to bactericidal UV light Purified water then flows to the mercury free ech2 o oxidation lamp Organ...

Страница 7: ...he recirculation icon this also refreshes the water quality parameters Product resistivity TOC Temperature There are several options to dispense ultrapure water Manual Dispensing To start dispensing i...

Страница 8: ...Manual Dispensing To start dispensing at low flow turn the dispense wheel one notch counter clockwise To start dispensing at full flow press down on the wheel Adjust flow rate Stop To increase the fl...

Страница 9: ...umetric dispensing a pre selected volume is auto dispensed and the last millilitres are distributed drop by drop for the user to manually stop the flow when the calibration mark is reached The assiste...

Страница 10: ...ze assign a name or experiment number In the user ID field enter the content of your choice max 15 characters 3 Export Scan the QR code from a mobile device to instantly retrieve the report Click on t...

Страница 11: ...nd better understand the root cause Advice is provided in the information section to help the user troubleshoot If the issue is not resolved select the Contact tab to find the technical support hotlin...

Страница 12: ...easily be reached in a few clicks or swipes as on a smart phone HOME SCREEN MAIN MENUS Home Water quality information Dispensing functions Alerts Alarms when applicable Dispense report when applicabl...

Страница 13: ...in a menu or app level screen 1 minute of inactivity generates a time out back to the home screen When in a pop up level screen there is no time out therefore the user remains on this screen until th...

Страница 14: ...iqconsumables com All cartridges have been designed to be easily replaced by a user A short cut is available to the replacement wizard by clicking on the replace pack button in the pop up Important ec...

Страница 15: ...sumable identification with card reading Some of the consumables come with a card for scanning and identification by the system For sys tem consumables the card must be scanned on the end of the dispe...

Страница 16: ...ture and TOC Also included is the total volume dispensed during the day Dispensing events report Each row is a dispense operation It is an archive of all the dispense reports To export data or preview...

Страница 17: ...ow maintenance alert Red technical issue alarm Production flow schematic Distribution flow schematic Troubleshooting Access a diagnostic summary This app contains all the necessary information in one...

Страница 18: ...quickly help users identify the right contact for any questions or issues Technical support hotline Application specialist Field service engineer System manager Maintenance operator In case of trouble...

Страница 19: ...ly Read only Periodic EDI rinsing period Read only Read only Tank configuration Tank capacity Read only Read only Tank refill setpoint Read only ech2o ASM lamp cycle start hour Read only Distribution...

Страница 20: ...old LC Pak VOC Pak EDS Pak LC Pak 0 1 m 0 1 m Pak Connectivity Local network Read only Date Time Lan guage Time zone Date Time Read only Select language Read only Password System manager Units Resisti...

Страница 21: ...ur Note By default this parameter is activated The default password is PASS If the system manager password is forgotten contact the technical support hotline Log off as a Manager 1 Press on the menu b...

Страница 22: ...at 00 01 on the selected date Safety flow stop A precautionary measure that stops a POD dispensing after having continuously delivered water for a certain duration Adjust to the preferred time by usi...

Страница 23: ...This can be personalized Click on the text box and input up to a maximum of 8 characters Screen brightness Adjust to the preferred brightness from 1 to 7 using the arrows or click on the box to access...

Страница 24: ...to increase the flow rate up until full flow is reached Release at the chosen flow rate 3 Press once to stop the flow dispensing Washer Distribution kit This accessory will allow distributing water fr...

Страница 25: ...particle concentration the inlet strainer clean frequency can be adapted Cartridge lifetimes Cartridges should be changed when the system alerts the user Saturated pack content will start to let ion a...

Страница 26: ...or a standardized way of presenting dates and times Date Calendar date representations are shown in the form 2018 11 06 YYYY MM DD YYYY indicates a four digit year MM indicates a two digit month of th...

Страница 27: ...e to the fact that these two parameters are not measured exactly at the same time TC2 The actual temperature compensated resistivity or conductivity values are displayed The TC2 setting should be used...

Страница 28: ...and effortless Available apps Note Dispensing in this menu is only available in the applicable wizards Maintenance screens Maintenance ongoing from the current facing POD Note To unlock all PODs from...

Страница 29: ...anta polishing cartridge IPAKQUAA1 1 or 2 IPAK Gard depending on system type Note 1 For IPAK Quanta before starting the rinse ensure there is at least 30L of water present If the IPAK Quanta rinse run...

Страница 30: ...ack installation twist until the click as described in the self guided wizard Pack Replacement Note The Milli Q IQ 7010 7015 systems have 3 packs 2 IPAK Gard 1 IPAK Quanta whereas the Milli Q IQ 7003...

Страница 31: ...Rinsing must be performed manually to complete the installation as indicated on the screen When completed press Finish to exit the wizard Note All certificates of quality are now available online Visi...

Страница 32: ...ct Care Cleaning app and click on Inlet strainer clean Follow the self guided wizards Cleaning the A10 TOC Monitor This is recommended when A new IPAK Quanta cartridge and IPAK Quanta ICP if applicabl...

Страница 33: ...e for further information PH Cleaning Follow the self guided wizard Contact your service representative for further information Depressurization This temporarily stops production and all water process...

Страница 34: ...e panels of the Milli Q system Hazardous voltage Hazard of electrical arc flash will cause death or serious injury Important Never open the system even when the power switch has been switched off It i...

Страница 35: ...be physically removed from its power source Installation and maintenance should only be done by a qualified person Appropriate personal protective equipment PPE must be worn and safe work practices m...

Страница 36: ...should be replaced when it appears to be clogged Tank is not re filled of pure water Open the drop down menu swiping down on the touchscreen to check the status of the pure production section Exit th...

Страница 37: ...ispense 1 Alarm with number of active alarm s Alert Home Menu Storage tank System manager logged in Mass storage USB connected Ethernet LAN status connected Slider ON Slider OFF Back Calendar entry Pr...

Страница 38: ...ndotoxins 0 001 EU mL with Biopak filter when installed and used in a laminar flow hood RNases 1 pg mL with Biopak filter DNases 5 pg mL with Biopak filter Proteases 0 15 g mL Flow rate 3 L h for IQ 7...

Страница 39: ...lter when installed and used in a laminar flow hood RNases 1 pg mL with Biopak filter DNases 5 pg mL with Biopak filter Proteases 0 15 g mL Flow rate up to 2 L min Volumetric Dispensing Accuracy 2 for...

Страница 40: ...iculates 0 1 m 1 particulates mL Bacteria 1 cfu mL Flow rate up to 2 L min Note These specifications are valid for a Milli Q IQ Element Unit fed by water produced by a Milli Q IQ 7003 7005 7010 7015 S...

Страница 41: ...requirements Parameter Value or range Pressure 1 bar P 6 bar Type of water Potable tap water Temperature 5 35 C 41 95 F Conductivity 2000 S cm 25 C Dissolved CO2 30 ppm Free Chlorine 3ppm Fouling Ind...

Страница 42: ...0 50 60 Hz 10 600 VA 100 240 V 10 Environmental Specific environmental conditions have to be respected to ensure normal operation of the Milli Q system Location indoor use only Ambient operating tempe...

Страница 43: ...he EU RED directive and or FCC regulations Standard EU We certify that these Lab Water Systems are designed and manufactured in application of the following European Council directives DIRECTIVE 2014...

Страница 44: ...Requirements and specifications 42 Dimensions and weights Water purification unit 7003 7005 Water purification unit 7010 7015...

Страница 45: ...Requirements and specifications 43 POD Storage tank 25L 50 L 100 L...

Страница 46: ...a wetted system with all its consumables but not any accessories Recycling Directive 2012 19 UE For European users only The symbol crossed bin on a product or its packaging indicates that the product...

Страница 47: ...ip ZFDISTPA1 Alarm relay cable ZMQ0ALCA1 Washer Distribution kit 230V Left ZWDK5L100 Washer Distribution kit 230V Right ZWDK5R100 Washer Distribution kit 115V Left ZWDK6L100 Washer Distribution kit 11...

Страница 48: ...h2 o ASM Lamp ASMUVLPA1 Ech2 o Bactericidal Lamp ZIXUVLPA1 Ech2 o oxidation lamp ZIQUVLPA1 Ech2 o A10 TOC Lamp ZFA10UVA1 Millipak 0 22 m filter MPGP002A1 Millipak Gold 0 22 m sterile filter MPGPG02A1...

Страница 49: ...l M Millipore Milli Q Q POD E POD Milli Q IQ Element A10 ech2 o IPAK Gard IPAK Quanta Millipak Biopak EDS Pak VOC Pak and LC Pak are trademarks of Merck KGaA Darmstadt Germany Millipore SAS is an affi...

Страница 50: ...powered OFF before plugging in or removing any components on the electronic board s The Milli Q IQ 7003 7005 7010 7015 has been tested by an independent and accredited company for compliance with EU...