Processes Application
Makeup process: The Tank refill setpoint can be modified, the RO recovery setpoint can
be decreased. The Tap feed pressure max. is given as information only.
ASM scheduling: If the option has been activated, ON time and cycle period can be
adjusted based on the system daily usage. It is not recommended to disable the ASM UV
lamp cycle.
Distribution process: Auto lab closed can be activated here. The auto lab closed must
be activated to refresh the tank water if the system was not in tank filling for more than 3
days and the tank level is above the tank refill setpoint.
The "Auto recirculation duration" can be adjusted from 15 to 60 min per hour, and the Post
recirculation from 30 to 60min per hour.
The "Tank flush setpoint" can be adjusted to the following range of value: 0-75%.
Options: The water sensor (or leak detector) can be activated here.
Measure outputs: Two measures can be selected from the list of parameters, these
parameters can be monitored by external devices.
Units: Set your desired units here.
Alarm and Alert Application
Alarm thresholds: Thresholds corresponding to the main system parameters can be
adjusted here. If a threshold has been crossed, the system will raise the corresponding
alert or alarm.
Example: RO Low TDS setting can be activated in some particular tap feed water cases.
If the tap feed water quality is too good, the RO membrane rejection performance maybe
poor and based on the system RO recovery, the system could raise alarms related to the
RO rejection, or tap feed conductivity. This parameter can be selected to suppress these
alarms.
External signals: This option must be installed and activated by your service
representative. Unlimited alerts or alarms can be selected from the list, these alarms can
be reported to an external device.
System Settings Application
LCD: Adjust the brightness of the display.
System Display | 33
Содержание HR 7060
Страница 1: ...Milli Q HX 7040 7150 HR 7060 7220 User Manual...
Страница 2: ......
Страница 9: ...Specifications Specifications 9...
Страница 12: ...System Dimensions HX and HR System dimensions 12 Specifications...
Страница 50: ...Ordering Information 50 Ordering Information...
Страница 54: ...Appendix 54 Appendix...