Sanitization and Cleaning Application allows cleaning software
wizards to be launched.
The RO membrane cleaning wizard will guide you through the different steps, indicate the
cleaning time and what is required in order to perform RO membrane cleaning. During the
first steps, until reagents are introduced into the system, you will be prompted by the
wizard, to validate or cancel the RO membrane cleaning.
When opening this application, two dates related to cleanings are shown:
• The "Performed" date indicates the last time a cleaning was done,
• The "Due date" indicates when the next cleaning is due.
Note: The cleaning timers can be adjusted in collaboration with your Qualified
Service representative.
The RO membrane CL
2
cleaning takes around 30 min and the RO membrane pH cleaning
takes around 75 min (see Maintenance Chapter for more information).
Tank emptying: Your tank will be emptied until it reaches the 0% level of water.
TOC cleaning: TOC cleaning process will be launched, the time duration is 1 hour.
System Display | 31