MAINTENANCE MENU

The maintenance of the system is easy thanks to the step-by-step instructions contained in the 

wizards. In addition, a new cartridge design means that pack removal and installation in the 

system unit is quick and effortless.

Available apps:

0

Note:

 Dispensing in this menu is only available in the applicable wizards.

Maintenance screens

 

Maintenance ongoing from the 

HMI.

 

Depressurization

This temporarily stops production and all 

water processes.

 

 

This function is not necessary during 

normal system operation. See

 self help 

guide section

.

Consumable installation wizards

A maintenance wizard can be accessed in different ways:

1. 

Through the Maintenance menu 

 

2. 

A shortcut from an Alert 

 or Alarm 

1

 pop-up  

3. 

A shortcut from a consumable pop-up

Install consumables

Care/cleaning

Depressurization

Содержание EQ 7000

Страница 1: ...User Manual Milli Q EQ 7000...

Страница 2: ...vity 15 Date Time Language 16 Units 17 INFORMATION MENU 18 System 18 History 18 Flow schematic 19 Troubleshooting 19 Contacts 19 MAINTENANCE MENU 20 Depressurization 20 Consumable installation wizards...

Страница 3: ...e The Milli Q EQ 7000 is intended to produce ultrapure water from a purified water source primarily for use in a variety of laboratories worldwide The product is designed to produce ultrapure water wi...

Страница 4: ...ed of ion exchange resins removing the bulk of the ions The purified water then flows to the UV 185nm oxidation lamp Organic molecules are partially oxidized resulting in ions These can then be captur...

Страница 5: ...ter within the system will recirculate for three minutes every hour This is to maintain a clean environment and to avoid contamination The system should not be turned off as this will stop the periodi...

Страница 6: ...ing 2 The system is powered on STEADY The water quality fits the requirements water quality values are below alarm setpoints BREATHING The system is in recirculation mode A water quality alarm is rais...

Страница 7: ...ispensing To do this press on the recirculation icon this also refreshes the water quality parameters Product resistivity TOC Temperature There are several options to dispense ultrapure water Manual D...

Страница 8: ...ed Water Quality values When the dispensing is finished the resistivity and temperature are instantly displayed on the left side of the HMI Ninety seconds later a new TOC value at the point of distrib...

Страница 9: ...pecific alert or alarm to get further information and better understand the root cause Advice is provided in the Information tab to help the user troubleshoot If the issue is not resolved select the C...

Страница 10: ...tions can easily be reached in a few clicks or swipes similar to a smartphone HOME SCREEN MAIN MENUS Home Water quality information Dispensing function Alerts Alarms when applicable Menu Consumables I...

Страница 11: ...l automatically appear To remove the keypad click on or on any zone outside the keypad Note When in a menu or app level screen 1 minute of inactivity generates a timeout back to the home screen When i...

Страница 12: ...information and a description of the purification technology Note Certificates of quality are now available online www mymilliqconsumables com All cartridges have been designed to be easily replaced b...

Страница 13: ...d only Feed quality measure Read only Alarm relay activation Read only Alarm relay configuration Custom Volume HMI configuration Brightness Buzzer Read only Alarm configuration Product resistivity Rea...

Страница 14: ...Note By default this parameter is activated The default password is PASS If the system manager password is forgotten contact the technical support hotline Log off as a Manager 1 Press on the menu butt...

Страница 15: ...15 minutes and is active This setting will not impact volumetric dispensing functions Screensaver This is the maximum duration of inactivity on the HMI before the screensaver starts 1 Activate deactiv...

Страница 16: ...f bad water quality is detected an alarm that stops water production will appear informing the user that the system is at risk and to check the feed water quality This will protect the downstream puri...

Страница 17: ...s should be changed when the system alerts the user Saturated pack will start to let ion and organic contaminants through This is called breakthrough and often starts with traces that cannot be detect...

Страница 18: ...ns are shown in the form 2020 07 06 YYYY MM DD YYYY indicates a four digit year MM indicates a two digit month of the year 01 through 12 DD indicates a two digit day of that month 01 through 31 Start...

Страница 19: ...rature compensated resistivity or conductivity due to the fact that these two parameters are not measured exactly at the same time TC2 The actual temperature compensated resistivity or conductivity va...

Страница 20: ...ng events report Each row is a dispense operation It is an archive of all the dispense reports To export data or preview a graph 1 Select a time period of interest with a start and end date 2 A graph...

Страница 21: ...ffective troubleshooting when in contact with the technical support hotline System information last service visit Water quality consumables installed dates Active alerts alarms autotest results Diagno...

Страница 22: ...able in the applicable wizards Maintenance screens Maintenance ongoing from the HMI Depressurization This temporarily stops production and all water processes This function is not necessary during nor...

Страница 23: ...artridges may not be fully rinsed The remaining rinse will need to be performed manually Important During the physical pack replacement in the water purification unit both cartridges must first be rem...

Страница 24: ...ion will ensure accuracy of your dispensing Note You will need a 1 0L graduated cylinder Select Care Cleaning app and click on Volumetric dispensing calibration Follow the self guided wizards UV lamp...

Страница 25: ...rom its power source Installation and maintenance should only be done by a qualified person Appropriate personal protective equipment PPE must be worn and safe work practices must be followed If the s...

Страница 26: ...ther cartridges installation IPAK other cartridges installation new cartridges not recognized Important During the physical replacement in the system unit both cartridges must first be removed before...

Страница 27: ...Dispensing Volumetric dispensing System manager logged in Start dispense Stop dispense 1 Alarm with number of active alarm s Alert Home Menu Storage tank Back Mass storage USB connected Ethernet LAN s...

Страница 28: ...installed and used in a laminar flow hood Pyrogens endotoxins 0 001 EU mL with Biopak filter RNases 2 pg mL with Biopak filter DNases 5 pg mL with Biopak filter Proteases 0 15 g mL with Biopak filter...

Страница 29: ...er supply converts mains voltage to 28 V The power supply is compatible worldwide Catalog Frequency Max power Voltage ZEQ7000T0 ZEQ7000TR 50 60 Hz 2Hz 170 VA 100 240 Vac 10 Environmental Specific envi...

Страница 30: ...d USB keys only work when formatted FAT32 NTFS format is not compatible Ethernet When connected through an Ethernet protocol the display interface can be accessed remotely using internet web browsers...

Страница 31: ...ssories Note For wall mounting installation use a concrete wall capable of support max load of 16 Kg Use 6mm diameter stud and fasteners capable to support max load of 4 Kg each Recycling Directive 20...

Страница 32: ...l ZEQ7C0NDC Internal Pressure Regulator ZF3000810 e Sure dispenser tip ZFDISTPA1 Alarm relay cable ZMQ0ALCA1 Consumables order at www mymilliqconsumables com Description Catalog Number IPAK Meta IPAK...

Страница 33: ...ore Milli Q Q POD IPAK Meta IPAK Quanta Jetpore IQnano Millipak Biopak EDS Pak VOC Pak and LC Pak are trademarks of Merck KGaA Darmstadt Germany Millipore SAS is an affiliate of Merck KGaA Darmstadt G...

Страница 34: ...r could exist This CAUTION sticker is used to refer to a surface that can be hot Disconnect and switch power off to allow surface to cool before servicing Hazard The Milli Q system must be connected t...

Отзывы: