HistoDream EW
MM164-001-Operator Manual
29
2.9. Reagents
Complete installation with reagents, by providing at least:
EMBEDDING REAGENTS:
5L of wax (5kg of solid wax); Milestone wax can be used:
REF 51439
SYNERGY WAX
Box including N. 6 x 2 kg pouch Synergy Wax, histology grade
2.10. Operating settings
When the System is switched on, the splash screen is displayed by default.
MILESTONE
VERSION X.X
To set the operating settings (language, temperature measurement unit, date format, auxiliary tweezers,
available dispenser and thermal module), turn off the dispensing module via the switch on the rear panel,
and then switch it back on.
While the splash screen is displayed, press the MANUAL MODE, UP and DOWN keys one after the other.
The following screen is displayed:
S.A.T.
MILESTONE v.x.x.
This message indicates that the unit is in programming mode.
After about 5 seconds the following screen opens:
SELECT LANGUAGE
ENGLISH
Use the UP or DOWN arrow keys to select the desired interface language.
Once selected, press the ENTER key to confirm.
TEMPER. SELECTION
CELSIUS
Use the UP or DOWN keys to select the CELSIUS or FAHRENHEIT
temperature units, and press ENTER to confirm the selection.
DATE FORMAT
EUROPEAN FORMAT
Use the UP or DOWN keys to select the date format:
-European format: DD-MM (day - month)
-American format: MM-DD (month - day)
Press ENTER to confirm
AUX. FORCEPS
ENABLED
Use the UP or DOWN arrow keys to select ENABLED if the optional
auxiliary tweezers are installed or DISABLED if they are not. Press ENTER
to confirm.
CONT. RESERVOIR HEAT
ENABLED
Use the UP or DOWN keys to select ENABLED and keep the following in
continuous operation mode:
-the wax cavity heating system
- the heating system of the tube to the dispenser
- the contact detector for dispensing wax
This option makes it possible to dispense wax at any time. Note that the
heating areas, tweezers and cold plate will not be operating.
Select DISABLED if the dispenser is not required to be continuously