2.7. General description
1. Wax cavity
2. Chamber for moulds (rear)
3. Chamber for cassettes (front)
4. Wax injection nozzle
5. Operating lever
6. Cold plate
7. Keyboard and display
8. Cooling point
9. Excess wax collection drawers
10. Optional heated tweezers connector
11. Tweezer holder holes
12. Wax removal grates
13. LED light
Wax cavity
Carefully close the wax tank cover.
• No recycled wax may be used in the HistoDream DM, Dispensing Module due to the contamination risk.
• Poor wax quality may lead to blockage. Use suitable wax.
• Topping up with different wax may cause breakages in wax blocks. It is recommended to always use the
same type of wax.
• Top up the wax carefully.
The cover must always be in the closed position, otherwise the adjusted temperature cannot be kept.
Hot surface
1
2
7
6
3
4
5
11
13
9
8
10
9
12