Configuration
EoSens GE Camera Manual
6 Configuration
The content of all Eo
Sens
registers is called a profile. There is space in non volatile memory
for 9 profiles: The PowerUpProfile, 4 user profiles and 4 factory profiles.
Any change of a specific register through the serial interface is immediately processed and
written to the volatile part of the memory and gets lost when power goes down. A command
must be used to store the actual setting in non volatile memory. After power-up the PowerUp-
Profile is loaded from the non-volatile to the volatile part of the memory.
A load or write command exchanges data between the CameraProfile and one of the four user
profiles. The four factory profiles can be read but not be written by any command. All values
are given in hexadecimal notation, e.g.: 0xff or 0ffh = 255.