LIST OF WARNINGS AND EVENTS
REV.0.01_COD.650184_A4
Page 46
Warning
Display message
Description
34
check temp.
Temperature difference between the two sensors is
greater than 1°C (1.8°F).
35
Serial connect.1
No connection between main PCB and loading
controller
36
Serial connect.2
No connection between the expansion board and the
keyboard (unloading side).
37
CANserialconnec.
No connection between master and slave (CANbus).
39
no tank heating
No heating in wash cabinet. No temperature increase
during allocated time.
40
no boilerheating
Problem with boiler heating.
41
no boilerheating
Problem with boiler heating.
42
magnetothermic
Steam heating fault
43
Phase sequence
Phase sequence is incorrect.
44
levels steamcond
Steam condenser fault
45
pump steam cond.
Min. and max. level active drain pump fault.
46
pump
Circulation pump on; pressure switch closed.
Circulation pump is rotating in the wrong direction.
47
flowmeter fail.1
DOS1 dispenser fault
48
flowmeter fail.2
DOS2 dispenser fault
49
flowmeter fail.3
DOS3 dispenser fault
50
flowmeter fail.4
DOS4 dispenser fault
51
flow.cold water
Water intake fault (cold water).
52
flow.warm water
Water intake fault (hot water).
53
flow.demi water
Water intake fault (fully demineralized water).
54
airfilter obstr.
Air filter is obstructed. Change the air filter.
Vacuum switch defective. No compressed air available
(check compressed air supply).
55
conduc. probe
Conductivity sensor fault.
56
conductivity
Conductivity exceeds maximum threshold.
58
no tank heating
No temperature increase during allocated time.
59
no boilerheating
No temperature increase during allocated time.
60
TIME
Maximum time exceeded in main wash phase
72
DRYING
Exhaust air pressure: fault when opening or closing
exhaust air valve.
73
DRYING
Exhaust air pressure: exhaust air limit switch (open and
close both active at the same time).
74
boilerprobelim°C
The boiler temperature is too high.
75
boiler probe
Boiler probe 2 defective
76
BOILER
The max. level on tank 3 is active and the min. level is
inactive.
77
BOILER
The max. level on tank 1 is active and the min. level is
inactive.
78
BOILER
The max. level on tank 2 is active and the min. level is
inactive.