OPERATION
REV.0.01_COD.650184_A4
Page 27
WARNING
The maximum load for each cycle is 35 kg (77 lb) (incl. basket).
Never operate the machine without a load carrier in place.
Before starting to use the machine, make sure that all the routine checks have been carried out.
Check the spray arm rotation.
The list below provides examples of load carriers and inserts which may be used when
reprocessing laboratory glassware and utensils.
Other accessories are available from Miele.
Upper basket with nozzles
Upper basket with spray arm
Lower basket with nozzles
Lower basket for pipettes
Insert for glassware
Insert for Petri dishes
Checks at the end of a program
Carry out a visual check of the wash items for cleanliness.
Are all lumened instruments still attached to the appropriate nozzles?
WARNING
Any lumened instruments which have become disconnected during
reprocessing must be reprocessed again.
Are the lumens free from obstructions?