Safety instructions and
warnings
,
These safety instructions and
warnings are in addition to those
found at the beginning of the
Installation and Operating
Instructions!
~
The H
2
O
2
solution must only be used
in special containers from "Dr. Weigert",
supplied by Miele, with the matching
adapters.
~
Always read and comply with the
material safety data sheets from the
manufacturers of the process
chemicals.
~
Exercise care when handling
H
2
O
2
solution - it is a corrosive
substance!
Always comply with all applicable
safety regulations!
Use protective eyewear and gloves.
~
Follow the manufacturer's
instructions for disposal of empty
containers.
~
The H
2
O
2
solution must never be
mixed with any other processing
chemicals. A violent chemical reaction,
such as a flash fire, may occur.
~
Only special tenside detergents
from "Dr. Weigert" should be used.
Connecting the H
2
O
2
container
The hose line for the H
2
O
2
container is
identified by a black label. It is supplied
without an adapter, because the
extraction systems for each container
differ by manufacturer.
^
Connect the appropriate adapter
from the manufacturer to the (black)
hose line.
^
Connect the H
2
O
2
container.
^
Start the
DOS2-FILL
service program.
In contrast to the other processing
chemicals, the H
2
O
2
container should
be completely empty before you
exchange it.
^
Wait until you see
Fill DOS 2
container
in the display before
installing a new container with H
2
O
2
-,
then start the
DOS2-FILL
service
program.
If the display shows
Check dispensing
system 2
, check the container and
dispensing line. The program
automatically pauses for this.
OXIVARIO and ORTHOVARIO Optional features
77