49
•
Printing the
STRING SUMMARY
results
1. Turn on the printer by pressing the
MODE
button.
2. Press the
ARROW
keys
to highlight the REPORTS icon, and press
ENTER
.
3. Press the
ARROW
keys
to highlight the STRING SUMMARY icon, and press
ENTER
.
4. Press the
ARROW
keys
to select the string name, and press
ENTER
.
5. Align the analyzer’s IR transmitter (to the right of the cable connector) with the IR receiver on
the printer (below the
MODE
button).
6. Press the
key. Keep the IR transmitter and receiver aligned for as long as it takes
to print.
7. Press the
BACK
key to end.
Printing the
JAR DATA
results
1. Turn on the printer by pressing the
MODE
button.
2. Press the
ARROW
keys
to highlight the REPORTS icon, and press
ENTER
.
3. Press the
ARROW
keys
to highlight the JAR DATA icon, and press
ENTER
.
4. Press the
ARROW
keys
to select the string name, and press
ENTER
.
5. Press the
ARROW
keys
to select a jar or strap screen, and press
ENTER
.
6. Align the analyzer’s IR transmitter (to the right of the cable connector) with the IR receiver on
the printer (below the
MODE
button).
7. Press the
key. Keep the IR transmitter and receiver aligned for as long as it takes
to print.
8. Repeat steps 5, 6, and 7 for each screen you want to print.
9. Press the
BACK
key to end.