43
•
===WARNING===
CONNECT TO
JAR XX
RED LEAD TO POSITIVE
POST, BLACK LEAD TO
NEGATIVE POST!
PRESS ‘5’ TO ADD JAR
NOTE:
XX is the battery ID of the battery whose place is being taken.
3. If you press
5
, the analyzer will add a placeholder to the stored data readings:
Battery XX
Voltage 0.0 volts
Conductance 0 mhos
When testing VOLTS THEN CONDUCTANCE, the PRESS 5 TO ADD JAR option is available only
during the voltage part of the test. During conductance testing the analyzer will skip batteries with
0 volts. You can add a placeholder by selecting the ADD BATTERY option in the REPORTS menu
at any time.
When testing is
fi
nished
You can add a placeholder later for a dead or missing battery via the ADD BATTERY option in the
REPORTS menu.
1. Press the
ARROW
keys
to highlight the REPORTS icon in the Main Menu, and press
ENTER
.
2. In the STRING INFO screen press the
ARROW
keys to highlight a string name.
3. In the test results screen, scroll to the battery after which you want the placeholder.
4. The analyzer will display:
INSERT PLACEHOLDER
0.0V and 0 MHOS
AT LOCATION
JAR XX
AND SHIFT OTHER DATA?
The place holder will be:
Battery ID 0
VOLTAGE 0.0 volts
Conductance 0 mhos
NOTE:
To add the placeholder at the end of the data list, select the TEST icon, press
ENTER
, and press the
5
key to add a jar.