IFU-001 Rev D June 2014
FDA 510(k) Cleared
Power
Frequency
(50/60 Hz)
Magnetic
Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment
Table 3: Guidance and manufacturer’s declaration
- electromagnetic immunity
The CP Relief Wand® is intended for use in the electromagnetic
environment specified below. The customer or user
should assure that it is used in such an environment.
Immunity
test
IEC 60501
Test level
Compliance
level
Electromagnetic environment
guidance
Portable and mobile RF
Communications equipment
should be used no closer to any
part of the device, including
cables, than the recommended
distance calculated from the
equation applicable to the
frequency of the transmitter’s
recommended separation distance
3 V/m
Radiated
RF IEC
61000-4-3
3 V/m
80 MHz to
2.5 GHz
Where “
P
” is the maximum output
power rating of the transmitter in
Watts (W) according to the
transmitter manufacturer and “d” is
the recommended separation distance
in meters “m”. Field strength from
fixed RF transmitters as determined
by the electromagnetic site survey, a.
should be less than the compliance
level in each frequency range. b.
interference may occur in the vicinity
of equipment marked with the
following symbol:
80 MHz to 800 MHz
d = 1.2 √
P
800 MHz to 2.5 GHz
d = 2.3 √
P