8.5.

CALIBRATION - Call

Before calibration the device has to be tared.

Reference weight has to be placed.

Allow to establish the value of the display.

Enters “CALIBRATION” mode with button 

“Enter/Prog”

 .

Use the arrows to correct the displayed value under placed reference weight.

Press the 

“Enter/Prog”

 button to confirm and output from calibration. 

Exit from “CALIBRATION” menu is performed automatically also 10 sec. after pressing a 
button and the changed value is stored.

8.6.

DOSES COUNTER – Count

 

The controller makes automatic reading of  number of switching of 

К3

 and stores them in 

nonvolatile memory. The counter is incremented by 1 when the measured weight exceeds the 
parameter 

doSE 3

. After reaching1000000 the counter is reset and continues its work. In the main 

menu there is an option to reset the counter (see “Operating guide” point 1)

8.7              SETTING THE DOSING PARAMETERS - Prog

Setting of parameters – mode 

“Prog”

. Changing the parameters is done through 

“Left”

 and  

“Up”

  buttons. The current digit for change is shown by flashing dot, using 

“UP”

 button the 

value is changed and using 

“Left”

 button the position of the flashing dot is changed. If  the 

change of sigh of the number is enabled it is changed when the flashing dot is on  leftmost 
digit

.

doSE1

Dose1 / weight of changing of 

K1

. Dimension and decimal point according to 

maximum measured weight and resolution of the sensor.

In mode A it represents a dose, and in mode B – level.

All features are in effect for the next two set-points.

doSE 2

Weight of changing of 

K2

doSE 3

Weight of changing of 

K3

 

After setting of 

“doSE 3”

“End“

 appears on the display. When pressing a button other than 

“Enter/Prog”

 proceed to re-setup of set-points. By pressing the 

“Enter/Prog”

 exit from 

setting mode.

8.8.

ERROR MESSAGE

Error

Function

“ErrorA” 

Measurement error. It is necessary the connection of measuring sensor with the 
device to be checked. Parameter 

ErLuvL

 determines the limit for activation of 

this message. 

“ErrorU” 

The tare weight together with some of operating set-points exceed the maximum
measured weight by the device. 

“tArEEr”

Tare weight >

 tArЕHi (% from HiL-t)

.

Document revision  2015-08-17

10

Содержание MS8357

Страница 1: ...COMPETENCE IN MEASUREMENT WEIGHT BATCHING CONTROLLER with Analog Output and MODBUS RTU communication MS8357 V 2 1 TECHNICAL DESCRIPTION AND USER MANUAL PLOVDIV 2015 Document revision 2015 08 17...

Страница 2: ...elay 3 A 250 VAC NO SPST 3 NPN OC iso 70 mA 70V Power supply 0 1 80 250 VAC 50 60 Hz 24 V DC Sensor supply sensitivity 1 5 VDC 1 7 mV V 2 Other Communication 0 No 1 RS485 MODBUS RTU Speed Conversion 1...

Страница 3: ...measurement speed up to 80 SPS Galvanic insulation of the measurement part from analogue output digital inputs and outputs and from RS485 Fast selection of up to 4 sets of parameters by combination o...

Страница 4: ...ime to Tare and Zero tmin 20 ms Digital outputs 3 K1 K2 K3 250 VAC 3 A SPST NO NPN OC iso 70 mA 70V transistor open collector unsulated from measurement part required an external power supply Unsulati...

Страница 5: ...wire connection the respective positive and negative terminals of the Exc and Ref is given short at side of the unit All sensors must be connected before the device is set tare and calibration Shield...

Страница 6: ...a termination resistor by Jumper R term next to the terminals of the RS485 VII OPERATING MODES The controller is used as a part of controlling of processes for dosing of materials Based on measured in...

Страница 7: ...switching Start 1 2 3 Stop can not be changed The controller gaps one or more stages if the weight at start exceeds their set points The whole process is limited in time by tOff parameter Start can be...

Страница 8: ...t less than dose 3 Its logic of action can be inverted by bit 2 of the ConFiG parameter Input Zero Start has function reset of the tare indications as after calibration In this mode the Start Stop but...

Страница 9: ...ital input The device can be tared by Tar input at any time except calibration procedure during setting the parameters and during filling for mode A This also applies to the input Zero Start clearing...

Страница 10: ...p buttons The current digit for change is shown by flashing dot using UP button the value is changed and using Left button the position of the flashing dot is changed If the change of sigh of the numb...

Страница 11: ...w 0 255 rS 485 Communication parameters For setup see below 0 255 Adr rS Modbus address 0 255 Flt Ar Average filter of the measured weight 1 1 5 Flt P Exponential filter of the measured weight 1 1 5 F...

Страница 12: ...128 Setting the configuration word Each bit in the configuration word has weight value and in the configuration word have to be entered sum of the weight values Example for setting of ConFiG paramete...

Страница 13: ...NICATION option The device is MODBUS RTU SLAVE with possibility for communication of 9600 bps by RS485 2WIRE line In standard performance without using a repeaters to one line can be connected up to 3...

Страница 14: ...net Doze3W4 16 Long Unsignet DPoint 102 LSByte Unsignet Doze2 18 Long Unsignet FLT J 103 Integer Unsignet Doze2W2 20 Long Unsignet FLT AR 104 MSByte Unsignet Doze2W3 22 Long Unsignet FLT EP 104 LSByte...

Страница 15: ...ted by the positions of both inputs for selecting of set point S1 and S2 The parameter A1 of the device screen is displayed x1024 ie read over the network 1 and the menu of the device would be 1024 Ca...

Страница 16: ...sport poor storage incorrect usage forces of nature failure to follow instructions and when others made an attempt to remove the defects In these cases the defect can only be removed for a fee Service...

Отзывы: