8.5.
CALIBRATION - Call
Before calibration the device has to be tared.
Reference weight has to be placed.
Allow to establish the value of the display.
Enters “CALIBRATION” mode with button
“Enter/Prog”
.
Use the arrows to correct the displayed value under placed reference weight.
Press the
“Enter/Prog”
button to confirm and output from calibration.
Exit from “CALIBRATION” menu is performed automatically also 10 sec. after pressing a
button and the changed value is stored.
8.6.
DOSES COUNTER – Count
The controller makes automatic reading of number of switching of
К3
and stores them in
nonvolatile memory. The counter is incremented by 1 when the measured weight exceeds the
parameter
doSE 3
. After reaching1000000 the counter is reset and continues its work. In the main
menu there is an option to reset the counter (see “Operating guide” point 1)
8.7 SETTING THE DOSING PARAMETERS - Prog
Setting of parameters – mode
“Prog”
. Changing the parameters is done through
“Left”
and
“Up”
buttons. The current digit for change is shown by flashing dot, using
“UP”
button the
value is changed and using
“Left”
button the position of the flashing dot is changed. If the
change of sigh of the number is enabled it is changed when the flashing dot is on leftmost
digit
.
doSE1
Dose1 / weight of changing of
K1
. Dimension and decimal point according to
maximum measured weight and resolution of the sensor.
In mode A it represents a dose, and in mode B – level.
All features are in effect for the next two set-points.
doSE 2
Weight of changing of
K2
doSE 3
Weight of changing of
K3
After setting of
“doSE 3”
,
“End“
appears on the display. When pressing a button other than
“Enter/Prog”
proceed to re-setup of set-points. By pressing the
“Enter/Prog”
exit from
setting mode.
8.8.
ERROR MESSAGE
Error
Function
“ErrorA”
Measurement error. It is necessary the connection of measuring sensor with the
device to be checked. Parameter
ErLuvL
determines the limit for activation of
this message.
“ErrorU”
The tare weight together with some of operating set-points exceed the maximum
measured weight by the device.
“tArEEr”
Tare weight >
tArЕHi (% from HiL-t)
.
Document revision 2015-08-17