GUIDANCE AND MANUFACTURER'S DECLARATION:
ELECTROMAGNETIC EMISSIONS
Emissions test
Compliance
Electromagnetic environment - guidance
Electromagnetic radiation dis-
turbance (radiated EMIS-
SIONS) CISPR 11
Group 1
The SpiderView uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
Disturbing voltage at supply ter-
minals conducted emissions
CISPR 11
Class B
The SpiderView is suitable for use in a pro-
fessional health care establishment.
Harmonic current EMISSIONS
IEC 61000-3-2
Non
applicable
Voltage variations, voltage fluc-
tuations and flicker IEC
61000-3-3
Non
applicable
GUIDANCE AND MANUFACTURER'S DECLARATION:
ELECTROMAGNETIC IMMUNITY
SpiderView is intended for use in the electromagnetic environment specified below. The
customer or the user of the SpiderView should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV con-
tact
±15 kV air
± 8 kV con-
tact
± 15 kV air
Home health care environment and profes-
sional health care establishment.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for
power sup-
ply lines
Non applica-
ble
Home health care environment and profes-
sional health care establishment.
± 1 kV for
input/output
lines
Surge
IEC 61000-4-5
± 1 kV be-
tween phas-
es
Non applica-
ble
Home health care environment and profes-
sional health care establishment.
± 2 kV be-
tween
line(s) to
earth
Magnetic field
at industrial rat-
ed frequency
(IEC61000-4-8)
30 A/m
Non applica-
ble
Home health care environment and profes-
sional health care establishment.
15.3.
15.4.
15. APPENDIX
36
SpiderView – UA10709B
Содержание SpiderView