― Holter recording
― High Resolution recording (HR)
― Holter rec HR recording
The Holter mode is always available if any flash card and cable is inserted into the device.
The High Resolution mode is only available if a 3-channel cable is inserted (XYZ shielded,
XYZ non shielded or standard 3 channels). The duration of the HR recording depends on the
size of the card.
The two can be performed sequentially: the recorder starts in HR mode and switches to Holter
mode after the programmed HR duration time.
Select one of the available modes using the arrow buttons and move to the next screen.
Holter mode
The Holter screen displays the setting of the most important parameters. According to the
flash card and the cable used, here you can select the recording quality and the recording
duration*.
* Duration longer than 24h available with SyneScope Holter software
Please refer to the table in this manual to check which combinations of modes and duration
you can select according to the size of your card. For patients weighing less than 10kg
(22lbs), Enhanced (ENH) or Uncompressed (NoComp) mode is recommended. By moving
the cursor to “REC” and pressing the "E" button you will start the Holter recording. A card
initialization phase will precede the recording.
WARNING
: Please check that the display of the recorder is switched off and that the LED
(6) is blinking before putting the recorder back into the pouch. This guarantees that the
initialization phase has finished and that the device has started to record correctly.
10.1.6.
10. STARTING THE HOLTER RECORDING
24
SpiderView – UA10709B