6.1 Manual cleaning
1) Separate the motor and Foot Switch connected in the Control Unit.
2) Prepare a cloth (preferably cotton) or soft brush moistened with isopropyl alcohol.
3) Clean foreign substance on the entire surface and in the gaps with cloth or brush
soaked in isopropyl alcohol for at least 2 minutes.
4) Repeat the cleaning process if foreign substance is found.
6.2 Inspection
1) Inspect whether there is any foreign substance visible to the naked eye.
2) Check the operating condition.
3) Repeat the cleaning process if foreign substance is found.
4) Repeat the above procedure for every single use.
Clean the product before and after use.
6.3 Sterilization
6.4 Change fuse
Sterilization of motor and irrigation tube by moist heat in accordance with ISO
17665-1, 2 in a steam sterilizer (Autoclave).
1) Gravity displacement type sterilizer
- At least 30 minutes at 121℃
- Drying time: 30 minutes.
2) Air-removal steam sterilization(pre-vacuum)
- At least 4 minutes at 132℃
- Drying time: 30 minutes.
Chapter 6. Maintenance
19/24
Extra fuse
Needed
to change
① Remove the fuse box with a flat screwdriver.
② Replace the fuses by the new ones.
③ Put the fuse box.