5.11 Auto Calibration
It is performed when the error of motor's rotation speed is higher than the set
speed during operation of implant engine.
When 'Set value adjustment button' are pressed at the same time,
the motor starts with a beep sound, the speed increases from 0 to 40,000 rpm
on the display, then the motor stops, and calibration is complete.
Check the set speed and actual speed by restarting the implant engine.
Auto Calibration is performed upon simultaneously
pressing both buttons longer than 2 seconds.
5.12 Operation of implant engine (Main function)
Implant engine operates upon completion of all settings.
When motor operates by pressing Foot switch, the letter 'R'
on display and the surrounding border flickers in turn.
The torque and speed indicate the current figures and
a beep sounds when altered torque value reaches 90%
of set value.
Motor stops when it reaches 100% of set value.
Control
Button
Chapter 55. Operation
LCD Display
18/24
5.10 Thread cutting function
Activating the thread cutting function is only possible with the
Tapping program.
① Forward rotation
② Reverse rotation
When loading
beyond the preset
torque limit value
③ Rotation stop
When foot switch off
④ Forward rotation
When foot switch on
LCD Display